Influence of eye tilt on corneal densitometry.

Ophthalmic Physiol Opt

Department of Mechanical, Materials and Aerospace Engineering, School of Engineering, University of Liverpool, Liverpool, UK.

Published: September 2022

Purpose: To investigate whether Pentacam densitometry readings are affected by corneal tilt.

Methods: In a prospective study, the right eyes of 86 healthy participants aged 42.8 ± 20.0 years (range 18-79 years) were imaged using Scheimpflug tomography. Elevation maps were exported to calculate corneal tilt using custom-made software, and densitometry readings were acquired directly from the corneal densitometry analysis add-on to the standard software Oculus Pentacam HR. Simple mediation analysis was applied to study age as a confounding factor in the correlation between corneal tilt and corneal densitometry.

Results: Corneal tilt and corneal densitometry are not independent from one another because age is significantly correlated with both corneal tilt (r = 0.50, p < 0.001) and corneal densitometry (r = 0.91, p < 0.001). Only 3.8% of the correlation between tilt and densitometry operates directly, while the remaining 96.2% depends on age.

Conclusions: Corneal tilt plays a role in corneal densitometry readings, even though the interaction is strongly influenced by age. Age is a well-known factor in densitometry readings that should be taken into consideration when interpreting Scheimpflug densitometry.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9543421PMC
http://dx.doi.org/10.1111/opo.13020DOI Listing

Publication Analysis

Top Keywords

corneal tilt
16
tilt corneal
12
corneal densitometry
12
corneal
9
densitometry readings
8
tilt
5
densitometry
5
influence eye
4
eye tilt
4
densitometry purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!