Objective: To determine whether more precise cup positioning can be achieved with robot-assisted total hip arthroplasty (THA) as compared to conventional THA.
Methods: In this study, between July 2019 and May 2021, 93 patients aged 23-75 years with osteonecrosis of the femoral head (ONFH) and adult developmental dysplasia of hip who underwent first hip surgery were included in the study. They were randomly assigned to either the robotic-assisted THA group (n = 45) or the conventional THA group (n = 48). After the operation, all patients were given routine rapid rehabilitation guidance. The duration of operation was recorded to estimate the learning curve through cumulative summation analysis. We compared the demographics, duration of operation, cup positioning, leg length discrepancy, hip offset, and Harris Hip Score between robot-assisted THA and manual THA. Precision in the positioning of the acetabular prosthesis using the MAKO system was also compared between the two groups.
Results: The mean duration of operation for the robot-assisted THA group was 91.37 ± 17.34 min (range: 63 to 135 min), which was significantly higher than that for the conventional THA group. When the number of procedures was increased to 13, the duration of operation in the robot-assisted group decreased significantly and gradually became stable. In terms of duration of operation, robot-assisted THA was associated with a learning curve of 13 cases. The mean amount of bleeding in the robot-assisted THA group was not significantly different from that in conventional THA group (328 ± 210 ml vs 315 ± 205 ml) (p = 0.741). There was no significant difference in the proportion of prostheses located within Lewinnek's safe zone between robot-assisted THA group and conventional THA group (69.81% vs 64.41%). The leg length discrepancy (LLD) was significantly smaller in the robot-assisted THA group than in the conventional THA group (p < 0.001), but both were within acceptable limits (10 mm). The inclination and anteversion angles of the acetabular prosthesis planned before operations were correlated with the actual measurement (r = 0.857 p < 0.001, r = 0.830, p < 0.001). After surgery, none of the patients experienced hip dislocation, aseptic loosening, or periprosthetic infection during the 3 months of follow-up.
Conclusion: The proportion of acetabular prostheses in the Lewinnek's safety zone was higher and the extent of LLD was significantly lower in the robot-assisted THA group, as compared to the same metrics in the conventional THA group. The MAKO robot improved the accuracy of implant placement in THA.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9251322 | PMC |
http://dx.doi.org/10.1111/os.13334 | DOI Listing |
Front Surg
January 2025
Rehabilitation Center, The First Rehabilitation Hospital in Shanghai, Shanghai, China.
Background: Telerehabilitation is gaining popularity in European and American countries, but whether it can be successfully implemented in China still lacks support from clinical studies.
Objective: This trial aimed to determine if a home-based telerehabilitation method is clinically noninferior to standard in-hospital face-to-face rehabilitation for elderly patients with total hip arthroplasty (THA) in China.
Methods: This multicenter randomized controlled trial was conducted from January 2021 to June 2022 at The First Rehabilitation Hospital in Shanghai, Shanghai Jiao Tong University affiliated Sixth People's Hospital and Shanghai Tongji University affiliated Tenth People's Hospital.
J Med Life
December 2024
Faculty of Medicine and Pharmacy, University of Oradea, Oradea, Romania.
This study aimed to assess pain intensity in patients who underwent total hip arthroplasty (THA) using the Visual Analogue Scale (VAS). The study was conducted between 2022 and 2023, enrolling 203 patients admitted to the Orthopedics Department. Participants self-reported their post-surgical pain 24 hours after THA by selecting a VAS rating that best represented their personal experience.
View Article and Find Full Text PDFOrthopadie (Heidelb)
January 2025
Endoprothesenregister Deutschland (EPRD), Berlin, Deutschland.
Objectives: To determine the influence of obesity on revision rates and mortality after primary elective hip and knee arthroplasty in Germany.
Materials And Methods: In the German Arthroplasty Registry (EPRD) there were 403,073 elective total hip arthroplasties (THA), 320,913 bicondylar total knee arthroplasties (TKA) and 48,480 unicondylar knee arthroplasties (UKA) with valid BMI available for analysis. Cumulative revision rates and 1‑year mortality was calculated for BMI groups.
J Arthroplasty
January 2025
Department of Orthopedic Surgery, NYU Langone Health, 301 E 17th St, New York, NY, USA, 10010. Electronic address:
Background: Despite advances in surgical techniques and postoperative prophylactic protocols, venous thromboembolism (VTE) events remain an important source of morbidity following total hip arthroplasty (THA). Prior research in cardiology and other surgical fields has suggested that statin medications may have a protective effect against VTE. Our study aimed to: 1) Assess if preoperative statin use was associated with decreased rates of VTE following THA, and 2) conduct a subgroup analysis of statin intensity and VTE events.
View Article and Find Full Text PDFThe current study aimed to objectively evaluate the fit of a rectangular, tapered stem to the severely dysplastic hips on the basis of the proximal femoral anatomy and the dimensional properties of the stem. It was hypothesized that the stem size planned with accordance to the diaphyseal canal width alone can accommodate the distal femur successfully with no sizing mismatch. Forty-six patients (53 hips) suffering from secondary osteoarthritis due to hip dysplasia scheduled for total hip arthroplasty (THA) with a subtrochanteric transverse shortening osteotomy were included.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!