EPI phase error correction with deep learning (PEC-DL) at 7 T.

Magn Reson Med

Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University, Shanghai, People's Republic of China.

Published: October 2022

Purpose: The phase mismatch between odd and even echoes in EPI causes Nyquist ghost artifacts. Existing ghost correction methods often suffer from severe residual artifacts and are ineffective with k-space undersampling data. This study proposed a deep learning-based method (PEC-DL) to correct phase errors for DWI at 7 Tesla.

Methods: The acquired k-space data were divided into 2 independent undersampled datasets according to their readout polarities. Then the proposed PEC-DL network reconstructed 2 ghost-free images using the undersampled data without calibration and navigator data. The network was trained with fully sampled images and applied to two- and fourfold accelerated data. Healthy volunteers and patients with Moyamoya disease were recruited to validate the efficacy of the PEC-DL method.

Results: The PEC-DL method was capable to mitigate the ghost artifacts in DWI in healthy volunteers as well as patients with Moyamoya disease. The fourfold accelerated results showed much less distortion in the lesions of the Moyamoya patient using high b-value DWI and the corresponding ADC maps. The ghost-to-signal ratios were significantly lower in PEC-DL images compared to conventional linear phase corrections, mini-entropy, and PEC-GRAPPA algorithms.

Conclusion: The proposed method can effectively eliminate ghost artifacts for full sampled and up to fourfold accelerated EPI data without calibration and navigator data.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.29317DOI Listing

Publication Analysis

Top Keywords

ghost artifacts
12
fourfold accelerated
12
data calibration
8
calibration navigator
8
navigator data
8
healthy volunteers
8
patients moyamoya
8
moyamoya disease
8
data
7
pec-dl
6

Similar Publications

Dual-polarity readout is a simple and robust way to mitigate Nyquist ghosting in diffusion-weighted echo-planar imaging but imposes doubled scan time. We here propose how dual-polarity readout can be implemented with little or no increase in scan time by exploiting an observed b-value dependence and signal averaging. The b-value dependence was confirmed in healthy volunteers with distinct ghosting at low b-values but of negligible magnitude at b = 1000 s/mm2.

View Article and Find Full Text PDF

LC-ESI-MS/MS is a preferred method for detecting and identifying metabolites, including those that are unpredictable from the genome, especially in basal metazoans like Cnidaria, which diverged earlier than bilaterians and whose metabolism is poorly understood. However, the unexpected appearance of a "ghost peak" for dopamine, which exhibited the same m/z value and MS/MS product ion spectrum during an analysis of Nematostella vectensis, a model cnidarian, complicated its accurate identification. Understanding the mechanism by which "ghost peaks" appear is crucial to accurately identify the monoamine repertoire in early animals so as to avoid misassignments.

View Article and Find Full Text PDF

Impact of uterine displacement on T-weighted image quality in the female pelvic MRI.

Radiol Phys Technol

November 2024

Department of Radiology and Nuclear Medicine, Fukushima Medical University School of Medicine, 1 Hikariga-Oka, Fukushima City, Fukushima, 960-1295, Japan.

Despite the importance of T-weighted image in clinical practice, artifacts can significantly degrade image quality and affect diagnosis. This study quantitatively analyzed uterine displacement and surveyed the relationship between the image quality of fast-spin-echo-T-weighted image of the female pelvis and quantitative value of uterine displacement. Overall, 147 women (mean age, 46.

View Article and Find Full Text PDF

Purpose: To evaluate the feasibility of a high-precision single-shot fast spin-echo (SS-FSE) sequence using the deep learning-based Precise IQ Engine (PIQE) algorithm in comparison with standard SS-FSE for T2-weighted MR imaging of the abdomen, and to compare the image quality with a multi-shot (MS)-FSE sequence using the PIQE algorithm.

Methods: This retrospective study included 105 patients who underwent abdominal MR including T2-weighted sequences using the PIQE reconstruction algorithm. The image quality, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in high-precision SS-FSE sequences using PIQE were compared to those in standard SS-FSE without PIQE and MS-FSE sequences using PIQE.

View Article and Find Full Text PDF

In this work, echo-planar spectroscopic imaging (EPSI) with flyback readout gradient-echo train was implemented in a preclinical MR scanner. The aim of this study is to visualize and quantify the ghost spectral lines produced by two, three and four interleaved echo trains with different amplitudes of the readout gradients, and to investigate the feasibility of the flyback data acquisition in micro-imaging of small animals. Applied multi-slice EPSI sequence utilizes asymmetric gradient-echo train that combines the shortest possible rewind gradients with readout gradients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!