Objective: The left heart bypass is currently the most frequent adjunct to provide distal aortic perfusion during aortic clamping. However, closed-circuits cannot respond to sudden hemodynamic fluctuations during aortic clamping which may lead to complications. In this report, we aim to give the technical aspects of reservoir-added centrifugal pump circuit system, its implementation and the clinical outcomes.

Methods: Between 2002 and 2020, the data of 35 patients underwent aortic aneurysm repair with the use of modified pump circuit were analyzed. The mean age was 53.4 years (range 24-73) and 91.4% of all was male. Preoperative demographics, intraoperative pump data and postoperative clinical outcomes were reported.

Results: Thoracoabdominal incision was used in 22 patients (62.9%) and cerebrospinal fluid (CSF) drainage catheter was placed in 15 patients (42.9%). The left inferior pulmonary vein for outflow and the left femoral artery for inflow cannulation were used preferably. Median duration of pump support was 50 mins (13-121) in the cohort. The median transfusion of red packed cells was 2 units. Renal failure was observed in two patients and permanent paraplegia in one patient. Only one patient died in the elective group (1/34) and one patient in the emergent.

Conclusions: The reservoir-added centrifugal pump is an effective, practical, and flexible perfusion system which should be in the armamentarium of surgeons in order to solve difficult problems during open repair of descending and thoracoabdominal aortic aneurysms.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15385744221108049DOI Listing

Publication Analysis

Top Keywords

centrifugal pump
12
pump circuit
12
thoracoabdominal aortic
8
aortic aneurysm
8
aneurysm repair
8
aortic clamping
8
reservoir-added centrifugal
8
pump
6
aortic
6
distal perfusion
4

Similar Publications

This study aimed to summarize our experience using a systematic approach to reduce blood transfusions in acute type A aortic dissection (ATAAD) surgery.From August 2016 to June 2020, 326 patients underwent ATAAD surgery in our center employing a systematic approach, which primarily included the following: Liu's aortic root repair technique, Liu's aortic arch inclusion technique with a frozen elephant trunk, moderate-to-mild hypothermia circulatory arrest, and application of centrifugal pump in cardiopulmonary bypass circuit. Patients were divided into two groups based on whether they had blood product transfusion during their hospital stay: transfusion group and transfusion-free group.

View Article and Find Full Text PDF

Nutritional Intake in Venovenous ECMO Patients: A Single-Center Study in a North American PICU.

Nutrients

November 2024

Department of Pediatrics and Human Development, College of Human Medicine, Michigan State University, Grand Rapids, MI 49503, USA.

Background/objectives: Multiple independent variables were analyzed to determine total nutritional intake (caloric and protein), while reviewing vasoactive inotrope scores (VISs).

Methods: Retrospective data were collected for nutritional intake (N = 64), daily VISs, extracorporeal membrane oxygenation (ECMO) complications, delays in nutritional intake (>48 h), reason for delay, and interruptions in nutrition support.

Results: VISs and ECMO characteristics analyzed by box-whisker plots demonstrated that ECMO complications prior to 24 h, cardiac arrest 24 h prior to ECMO, pediatric ECMO patients, venoarterial ECMO type, having a cardiac ECMO indication, and ECMO centrifugal pump had higher VISs.

View Article and Find Full Text PDF

A comprehensive analysis of in vitro pumps used in cardiovascular research is provided in this review, with a focus on the characteristics of generated flows and principles of flow generations. The cardiovascular system, vital for nutrient circulation and waste removal, generates complex hemodynamics critical for endothelial cell function. Cardiovascular diseases (CVDs) could be caused by the disturbances in these flows, including aneurysms, atherosclerosis, and heart defects.

View Article and Find Full Text PDF

Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates.

Artif Organs

November 2024

Department of Cardiovascular and Thoracic Surgery, Institute of Cardiology, Pitié Salpêtrière Hospital, Assistance Publique Hôpitaux de Paris (AP-HP), Sorbonne University, Paris, France.

Background: An innovative solution has been recently proposed for the treatment of heart failure with preserved ejection fraction (HFpEF), using a centrifugal mechanical circulatory support (MCS) device. We sought firstly to assess the hemocompatibility of the proposed device. HFpEF treatment requires the blood pump to operate at low blood flow rate (0.

View Article and Find Full Text PDF

Blood pumps, critical components in ventricular assist devices and extracorporeal membrane oxygenation systems, are primarily evaluated based on their ability to minimize blood damage through optimized design. Despite extensive research, the impact of impeller blade thickness and the proximity of rotating and stationary surfaces remains insufficiently explored. This study presents a comprehensive analysis, combining experimentally validated numerical simulations with an advanced Lagrangian approach, to compare the hemodynamic and hemolytic performance of three centrifugal pumps.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!