Background And Purpose: Spinal stereotactic ablative body radiotherapy (SABR) requires high precision. We evaluate the intrafraction motion during cone-beam computed tomography (CBCT) guided SABR with different immobilization techniques.

Material And Methods: Fifty-seven consecutive patients were treated for 62 spinal lesions with SABR with positioning corrected in six degrees of freedom. A surface monitoring system was used for patient set up and to ensure patient immobilization in 65% of patients. Intrafractional motion was defined as the difference between the last CBCT before the start of treatment and the first CT afterwards.

Results: For all 194 fractions, the mean intrafractional motion was 0.1 cm (0-1.1 cm) in vertical direction, 0.1 cm (0-1.1 cm) in longitudinal direction and 0.1 cm (0-0.5 cm) in lateral direction. A mean pitch of 0.6° (0-4.3°), a roll of 0.5° (0-3.4°) and a rotational motion of 0.4° (0-3.9°) was observed. 95.5% of the translational errors and 95.4% of the rotational errors were within safety range. There was a significantly higher rotational motion for patients with arms along the body (p = 0.01) and without the use of the body mask (p = 0.05). For cervical locations a higher rotational motion was seen, although not significant (p = 0.1). The acquisition of an extra CBCT was correlated with a higher rotational (pitch) motion (p = 0 < 0.01).

Conclusion: Very high precision in CBCT guided and surface-guided spinal SABR was observed in this cohort. The lowest intrafraction motion was seen in patients treated with arms above their head and a body mask. The use of IGRT with surface monitoring is an added value for patient monitoring leading to treatment interruption if necessary.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9172170PMC
http://dx.doi.org/10.1016/j.phro.2022.04.006DOI Listing

Publication Analysis

Top Keywords

rotational motion
12
higher rotational
12
spinal stereotactic
8
stereotactic ablative
8
intrafractional motion
8
01 cm 0-11 cm
8
direction 01 cm
8
motion
7
rotational
5
precision image-guided
4

Similar Publications

Establishing normative values and understanding how proprioception varies among body parts is crucial. However, the variability across individuals, especially adolescents, makes it difficult to establish norms. This prevents further investigation into classifying patients with abnormal proprioception.

View Article and Find Full Text PDF

The humeral head is the second most common anatomical site of osteonecrosis after the femoral head. Studies have reported satisfactory clinical outcomes after shoulder arthroplasty to treat osteonecrosis of the humeral head (ONHH). However, there are concerns regarding implant longevity in relatively young patients.

View Article and Find Full Text PDF

A novel air-to-liquid mass transfer system using wetted rotating membranes was designed to enhance air-to-liquid carbon dioxide (CO) mass transfer efficiency. Traditional methods, such as sparging, are energy-intensive, but the rotating membrane reduces energy demands by optimising membrane wetting via rotational motion. Experimental tests were conducted using a small-scale system with a membrane width of 0.

View Article and Find Full Text PDF

Purpose: This study aimed to explore the potential application of NAO in guiding patients through rehabilitative exercises using external audiovisual stimuli, focusing on temporospatial control in terms of range of motion (ROM), execution time and movement smoothness.

Methods: This is a preliminary analysis involving ten healthy volunteers and two patients with shoulder musculoskeletal disorders. The protocol was developed in two phases (III and IV) with different ROM limits and including flexion-extension (FE), external-rotation (ER) and internal-rotation (IR) exercises, performed at two speeds and both with and without NAO assistance.

View Article and Find Full Text PDF

Background: Traditional examinations of anterior cruciate ligament (ACL) injuries focus primarily on static assessments and lack the ability to evaluate dynamic knee stability. Hence, a dynamic scoring system for knee function is needed in clinical settings. This study aimed to propose a dynamic scoring system based on a large sample of normative six-degree-of-freedom (6-DOF) knee kinematics during gait, and validate its correlation with conventional outcome measurements in assessing ACL-injured knees.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!