Accuracy of dental implant surgery using dynamic navigation and robotic systems: An in vitro study.

J Dent

Department of Second Dental Center, Shanghai Ninth People's Hospital, College of Stomatology, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Shanghai Key Laboratory of Stomatology, Research Unit of Oral and Maxillofacial Regenerative Medicine, Chinese Academy of Medical Sciences, Shanghai Jiao Tong University, Shanghai Jiao Tong University School of Medicine, 280 Mohe Road, Shanghai, China. Electronic address:

Published: August 2022

Objectives: To compare the accuracy of dental implant placement using a dynamic navigation and a robotic system.

Methods: Eighty three-dimensional (3D) printed phantoms, including edentulous and partially edentulous jaws, were assigned to two groups: a dynamic navigation system (Beidou-SNS) group and a robotic system (Hybrid Robotic System for Dental Implant Surgery, HRS-DIS) group. The entry, exit and angle deviations of the implants in 3D world were measured after pre-operative plans and postoperative cone-beam computed tomography (CBCT) fusion. A linear mixed model with a random intercept was applied, and a p value <.05 was considered statistically significant.

Results: A total of 480 implants were placed in 80 phantoms. The comparison deviation of the dynamic navigation system and robotic system groups showed a mean (± SD) entry deviation of 0.96 ± 0.57 mm vs. 0.83 ± 0.55 mm (p=0.04), a mean exit deviation of 1.06 ± 0.59 mm vs. 0.91 ± 0.56 mm (p=0.04), and a mean angle deviation of 2.41± 1.42° vs. 1 ± 0.48° (p<0.00).

Conclusions: The implant positioning accuracy of the robotic system was superior to that of the dynamic navigation system, suggesting that this prototype robotic system (HRS-DIS) could be a promising tool in dental implant surgery.

Clinical Significance: This in vitro study is of clinical interest because it preliminarily shows that a robotic system exhibits lower deviations of dental implants than a dynamic navigation system, in dental implant surgery, in both partially and completely edentulous jaws. Further clinical studies are needed to evaluate the current results.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jdent.2022.104170DOI Listing

Publication Analysis

Top Keywords

dental implant
12
dynamic navigation
12
accuracy dental
8
implant surgery
8
navigation robotic
8
robotic system
8
surgery dynamic
4
robotic
4
robotic systems
4
systems vitro
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!