  • Implantable brain stimulation devices are being developed to treat brain conditions, requiring complex validation due to their interaction with patients and environmental factors.
  • A hardware-in-the-loop (HIL) testing framework has been introduced to simulate neural dynamics for evaluating adaptive deep brain stimulation (DBS) devices before moving to animal or human trials.
  • The study illustrates how real-time computational models were used to assess the effectiveness of adaptive DBS algorithms for managing epileptic seizures and Parkinson's disease symptoms using the new testing framework.

Article Abstract

Implantable brain stimulation devices continue to be developed to treat and monitor brain conditions. As the complexity of these devices grows to include adaptive neuromodulation therapy, validating the operation and verifying the correctness of these systems becomes more complicated. The new complexities lie in the functioning of the device being dependent on the interaction with the patient and environmental factors such as noise and artifacts. Here, we present a hardware-in-the-loop (HIL) testing framework that employs computational models of pathological neural dynamics to test adaptive deep brain stimulation (DBS) devices prior to animal or human testing. A brain stimulation and recording electrode array is placed in the saline tank and connected to an adaptive neuromodulation system that measures and processes the synthetic signals and delivers stimulation back into the saline tank. A data acquisition system is used to detect the stimulation and provide feedback to the computational model in order to simulate the effects of stimulation on the neural dynamics. In this study, we used real-time computational models to emulate the dynamics of epileptic seizures observed in the anterior nucleus of the thalamus (ANT) in epilepsy patients and beta band (11-35 Hz) oscillations observed in the subthalamic nucleus (STN) of Parkinson's disease (PD) patients. These models simulated neuronal responses to electrical stimulation pulses and the saline tank tested hardware interactions between the detection algorithms and stimulation interference. We tested and validated the operation of adaptive DBS algorithms for seizure and beta band power suppression embedded in an implantable DBS system (Medtronic Summit RC+S). This study highlights the utility of the proposed hardware-in-the-loop framework to systematically test the adaptive DBS systems in the presence of system aggressors such as environmental noise and stimulation-induced electrical artifacts. This testing procedure can help ensure correctness and robustness of adaptive DBS devices prior to animal and human testing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9125865PMC
http://dx.doi.org/10.1115/1.4054083DOI Listing

Publication Analysis

Top Keywords

computational models
12
brain stimulation
12
saline tank
12
adaptive dbs
12
stimulation
8
adaptive neuromodulation
8
neural dynamics
8
test adaptive
8
dbs devices
8
devices prior
8

Similar Publications

Background: Hip morphology variations, particularly in femoral neck shaft angle (NSA) and iliac wing width (IWW), have been associated with gluteal tendinopathy. However, the biomechanical implications of these morphological differences on gluteal muscle function are not well understood. This study investigates how NSA and IWW influence gluteal muscle forces, moment arms, and estimated tendon loads during walking, aiming to provide insights into the potential biomechanical pathways that may contribute to altered lateral hip loading patterns.

View Article and Find Full Text PDF

Background: This study aims to quantify intratumoral heterogeneity (ITH) using preoperative CT image and evaluate its ability to predict pathological high-grade patterns, specifically micropapillary and/or solid components (MP/S), in patients diagnosed with clinical stage I solid lung adenocarcinoma (LADC).

Methods: In this retrospective study, we enrolled 457 patients who were postoperatively diagnosed with clinical stage I solid LADC from two medical centers, assigning them to either a training set (n = 304) or a test set (n = 153). Sub-regions within the tumor were identified using the K-means method.

View Article and Find Full Text PDF

Adolescence is a developmental period of relative volatility, where the individual experiences significant changes to their physical and social environment. The ability to adapt to the volatility of one's surroundings is an important cognitive ability, particularly while foraging, a near-ubiquitous behaviour across the animal kingdom. As adolescents experience more volatility in their surroundings, we predicted that this age group would be more adept than adults at using exploration to adjust to volatility.

View Article and Find Full Text PDF

Automated stenosis estimation of coronary angiographies using end-to-end learning.

Int J Cardiovasc Imaging

January 2025

Novo Nordisk Foundation Center for Protein Research, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.

The initial evaluation of stenosis during coronary angiography is typically performed by visual assessment. Visual assessment has limited accuracy compared to fractional flow reserve and quantitative coronary angiography, which are more time-consuming and costly. Applying deep learning might yield a faster and more accurate stenosis assessment.

View Article and Find Full Text PDF

Hypophosphatasia (HPP) is a congenital bone disease caused by tissue-nonspecific mutations in the alkaline phosphatase gene. It is classified into six types: severe perinatal, benign prenatal, infantile, pediatric, adult, and odonto. HPP with femoral hypoplasia on fetal ultrasonography, seizures, or early loss of primary teeth can be easily diagnosed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!