4D Flow MRI is a diagnostic tool that can visualize and quantify patient-specific hemodynamics and help interventionalists optimize treatment strategies for repairing coarctation of the aorta (COA). Despite recent developments in 4D Flow MRI, shortcomings include phase-offset errors, limited spatiotemporal resolution, aliasing, inaccuracies due to slow aneurysmal flows, and distortion of images due to metallic artifact from vascular stents. To address these limitations, we developed a framework utilizing Computational Fluid Dynamics (CFD) with Adaptive Mesh Refinement (AMR) that enhances 4D Flow MRI visualization/quantification. We applied this framework to five pediatric patients with COA, providing in-vivo and in-silico datasets, pre- and post-intervention. These two data sets were compared and showed that CFD flow rates were within 9.6% of 4D Flow MRI, which is within a clinically acceptable range. CFD simulated slow aneurysmal flow, which MRI failed to capture due to high relative velocity encoding (V). CFD successfully predicted in-stent blood flow, which was not visible in the in-vivo data due to susceptibility artifact. AMR improved spatial resolution by factors of 10 to 10 and temporal resolution four-fold. This computational framework has strong potential to optimize visualization/quantification of aneurysmal and in-stent flows, improve spatiotemporal resolution, and assess hemodynamic efficiency post-COA treatment.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11034844 | PMC |
http://dx.doi.org/10.1007/s10439-022-02980-7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!