Background And Aims: Endoscopic ultrasound (EUS) guided choledochoduodenostomy using a lumen-apposing metal stent (LAMS) allows access to the biliary system with a sufficiently large diameter stent. However, an appropriate endoscope for this purpose is required. We evaluated the feasibility of per-oral cholangioscopy (POC) using a multibending ultra-slim endoscope through a LAMS for EUS-guided choledochoduodenostomy.

Methods: Twelve patients who underwent EUS-guided choledochoduodenostomy, followed by POC via a LAMS, were enrolled. POC was performed with an multibending ultra-slim endoscope. The primary outcome was technical success, defined as examination of both hepatic ducts and either the distal common bile duct or stricture site within 10 min. The types of intervention and adverse events were also assessed.

Results: Technical success was achieved in all cases with the multibending ultra-slim endoscope. Narrow-band imaging endoscopy was performed in five cases (41.7%) and POC-guided targeted biopsy was performed in three cases (25.0%). Stone extraction during POC was performed in two patients (16.7%) and foreign body removal from the intrahepatic duct was performed in one patient. One patient underwent additional metal stent insertion under direct visualization. After POC, no stent migration or severe adverse events were seen in any case.

Conclusions: POC using a multibending ultra-slim endoscope for diagnostic or therapeutic purposes can be performed effectively and safely through a LAMS after EUS-guided choledochoduodenostomy.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10620-022-07531-7DOI Listing

Publication Analysis

Top Keywords

multibending ultra-slim
20
ultra-slim endoscope
20
metal stent
12
lumen-apposing metal
8
poc multibending
8
lams eus-guided
8
eus-guided choledochoduodenostomy
8
poc performed
8
technical success
8
adverse events
8

Similar Publications

Background/purpose: Compared with currently available duodenoscopy-assisted systems, direct peroral cholangioscopy (DPOC) using an ultra-slim endoscope is limited by technical difficulties. The multibending (MB) ultra-slim endoscope was introduced as a dedicated cholangioscope for DPOC to challenge the technical problem. We retrospectively analyzed the clinical utility of DPOC using an MB endoscope with free-hand insertion into the bile duct in patients with biliary diseases.

View Article and Find Full Text PDF

Background And Aims: Endoscopic ultrasound (EUS) guided choledochoduodenostomy using a lumen-apposing metal stent (LAMS) allows access to the biliary system with a sufficiently large diameter stent. However, an appropriate endoscope for this purpose is required. We evaluated the feasibility of per-oral cholangioscopy (POC) using a multibending ultra-slim endoscope through a LAMS for EUS-guided choledochoduodenostomy.

View Article and Find Full Text PDF

Background And Aims: Currently available peroral cholangioscopy (POC) is a duodenoscopy-assisted procedure that does not involve directly inserting an endoscope into the biliary tree. A prototype multibending (MB) ultra-slim endoscope has been developed as a dedicated cholangioscope to overcome the technical difficulties of direct POC. In this study, we evaluated the efficacy of the new MB ultra-slim endoscope compared with a conventional ultra-slim endoscope for free-hand insertion of an endoscope into the bile duct for direct POC without the assistance of accessories.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!