Purpose: To measure net corneal astigmatism using optical coherence tomography (OCT) (Avanti) in individuals with keratoconus and compare the repeatability and accuracy with those obtained using Scheimpflug imaging (Pentacam HR).

Setting: Casey Eye Institute, Portland, Oregon.

Design: Prospective cross-sectional observational study.

Methods: Net corneal astigmatism was calculated in keratoconic and normal eyes using OCT and Scheimpflug imaging with 4 settings-pupil or vertex centration settings with a 3 or 4 mm circular analytical zone. Corneal elevation maps were obtained from OCT images and fitted with the Zernike polynomials to obtain net corneal astigmatism. Manifest refraction astigmatism was used to evaluate the accuracy of net corneal astigmatism measurements. The coefficient of repeatability from 2 repeated measures was calculated.

Results: 46 eyes with manifest or subclinical keratoconus and 52 normal control eyes were analyzed. For OCT measurements in keratoconus, better accuracy was achieved with pupil centration and 3 mm analytical zone; however, better repeatability was achieved with vertex centration and 4 mm analytical zone (coefficient of repeatability = 0.53 diopters, the Fligner-Killeen test with Bonferroni adjustment P < .0017). Agreement with manifest refraction was significantly better with OCT compared with that using Pentacam HR (generalized mixed-effect model with Bonferroni adjustment P < .00625). No statistically significant difference was found between instruments or settings in control eyes.

Conclusions: OCT was able to measure net corneal astigmatism with better accuracy and precision in keratoconic eyes than the Pentacam HR. Measurements may be more accurate using pupil centration and a smaller analytical zone in patients with keratoconus.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10414777PMC
http://dx.doi.org/10.1097/j.jcrs.0000000000000977DOI Listing

Publication Analysis

Top Keywords

corneal astigmatism
24
net corneal
20
analytical zone
16
measure net
8
scheimpflug imaging
8
vertex centration
8
manifest refraction
8
coefficient repeatability
8
better accuracy
8
pupil centration
8

Similar Publications

Distinct Ocular Surface Microbiome in Keratoconus Patients Correlate With Local Immune Dysregulation.

Invest Ophthalmol Vis Sci

January 2025

GROW Research Laboratory, Narayana Netralaya Foundation, Bangalore, India.

Purpose: Keratoconus (KC) is characterized by irregular astigmatism along with corneal stromal weakness and is associated with altered immune status. Tissue resident microbiomes are known to influence the immune status in other organs, but such a nexus has not been described in ocular conditions. Therefore, we examined the ocular surface microbiome of patients with KC and correlated it to the immune cell and tear molecular factor profiles.

View Article and Find Full Text PDF

Purpose: To determine the performance of TOTAL30 for Astigmatism (T30fA; Alcon; Fort Worth, TX, USA) contact lenses (CLs) in existing CL wearers who are also frequent digital device users.

Methods: This 1-month, 3-visit study recruited adult, 18- to 40-year-old subjects who were required to use daily digital devices for at least 8 hours per day. All subjects were refit into T30fA CLs.

View Article and Find Full Text PDF

Baseline and outcome stereoacuity of children with anisometropic amblyopia undergoing dichoptic amblyopia treatment.

J AAPOS

January 2025

Pediatric Vision Laboratory, Retina Foundation of the Southwest, Dallas, Texas; Optometry & Vision Science, University of Waterloo, Waterloo, Ontario, Canada.

Background: One rationale for dichoptic amblyopia therapy is that it may promote recovery of binocular function. Yet data on binocular outcomes in anisometropic amblyopia following dichoptic therapy are sparse. We report factors associated with pre- and post-treatment binocular function in anisometropic amblyopia, and examine binocular function in children who recover normal visual acuity compared to those with residual amblyopia.

View Article and Find Full Text PDF

Purpose To determine the corneal topographic characteristics of children with blepharokeratoconjunctivitis (BKC), a chronic inflammatory ocular surface disease involving the lid margin, conjunctiva, and cornea. Methods The corneal topography of 21 children diagnosed with BKC between March 2008 and June 2019 at a single institution were reviewed retrospectively. Pachymetry and asymmetry indices were also analyzed.

View Article and Find Full Text PDF

Background: Strabismus is a potential cause of ocular morbidity.

Objective: The aim of this study was to determine the frequency, types of manifest strabismus and co-morbidities among patients attending a referral paediatric ophthalmology and strabismus clinic in Calabar, Nigeria.

Methods: A retrospective review of case-notes of patients attending the paediatric ophthalmology and strabismus clinic from January 1, 2017 to December 31, 2019 was done.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!