The thermodilution (TD) method is routinely used for the estimation of cardiac output (Q̇). However, its accuracy, compared with the gold-standard Fick method, where systemic oxygen uptake (V̇O) is directly measured, and Q̇ calculated from V̇O and the arterio-venous oxygen difference ("direct" Fick), has not been well validated. The present study determined the agreement between TD and Fick methods in consecutive patients who underwent pulmonary artery catheterization for a broad range of clinical conditions. This is a subanalysis of a previous study comparing the indirect versus Fick method based on a prospective, consecutive patient registry of 253 patients who underwent pulmonary artery catheterization for clinical indications at a single center between 1999 and 2005. We included patients that had an estimation of Q̇ both by the Fick method using measured V̇O by exhaled gas analyses from timed Douglas bag collections and by TD. Cardiac index was classified as low when ≤2.2 L/min/m or normal when >2.2 L/min/m. The median (25th, 75th percentile) age of the cohort was 59 (50,67) years, and 50% were female. A total of 43.5% had normal left ventricular function by ventriculography, and 25.7% had ischemic heart disease. Median overall Fick and TD Q̇ were 4.4 (3.5, 5.5) and 4.3 (3.7, 5.2) L/min, respectively (p = 0.04). The median absolute percent error between Fick and TD Q̇ was 17.5 (7.7, 28.4)%, with a typical error of 0.88 L/min (95% confidence interval [CI] 0.82 to 0.95). Median absolute percent error was comparable in the low (n = 118) and normal Q̇ (n = 135) groups (16.9% vs 18.9%, respectively, p = 0.88). typical error was 0.3 (95% CI 0.27 to 0.33) and 0.49 (95% CI 0.45 to 0.55) L/min/m in that comparison. Percent error >25% between Fick and TD Q̇ was observed in over 30% of patients. Overall, Fick and TD Q̇ modestly correlated (R = 0.64, p <0.001), with a nondirectional error introduced by TD Q̇ [mean bias of 0.21 (-2.2, 2.7) L/min]. There was poor agreement between TD and the gold-standard Fick method, highlighting the limitations of making clinical decisions based on TD.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9648100PMC
http://dx.doi.org/10.1016/j.amjcard.2022.04.026DOI Listing

Publication Analysis

Top Keywords

fick method
16
fick q̇
16
percent error
12
fick
10
estimation cardiac
8
cardiac output
8
versus fick
8
method measured
8
oxygen uptake
8
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!