Background: The COVID-19 pandemic has resulted in increased admissions with respiratory failure and there have been reports of oxygen failure and shortages of machines to deliver ventilation and Continuous Positive Airway Pressure (CPAP). Domiciliary ventilators which entrain room air have been widely used during the pandemic. Poor outcomes reported with non-invasive respiratory support using ventilators which lack an oxygen blender could be related to an unreliable Fraction of inspired O (FiO). Additionally, with concerns about oxygen failure, the variety of ventilator circuits used as well as differing peak inspiratory flow rates (PIFR) could impact on the FiO delivered during therapy with domiciliary ventilators.

Methods: In a series of bench tests, we tested the effect of choice of circuit and different PIFR on the FiO achieved during simulation of ventilation and CPAP therapy using domiciliary ventilators.

Results: FiO was highly dependent upon the type of circuit used with circuits with an active exhalation valve achieving similar FiO at lower oxygen flow rates than circuits using an exhalation port. During CPAP therapy, high PIFR resulted in significantly lower FiO than low PIFR.

Conclusions: This study has implications for oxygen usage as well as delivery of non-invasive respiratory support during therapy with domiciliary ventilators when these are used during the second wave of COVID-19.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7750247PMC
http://dx.doi.org/10.1177/1751143720980280DOI Listing

Publication Analysis

Top Keywords

non-invasive respiratory
12
respiratory support
12
domiciliary ventilators
12
therapy domiciliary
12
peak inspiratory
8
inspiratory flow
8
covid-19 pandemic
8
oxygen failure
8
flow rates
8
cpap therapy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!