Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test.

Braz J Cardiovasc Surg

Division of Cardiac Surgery, Escola Paulista de Medicina, Universidade Federal de São Paulo, São Paulo, São Paulo, Brazil.

Published: May 2022

AI Article Synopsis

Article Abstract

Introduction: Aortic valve bioprostheses ring fracture in valve-in-valve procedures has shown low complication rates and presents as an option in the treatment of patients at high risk for conventional surgery, avoiding high transvalvular gradients, which are associated with increased mortality. Some prostheses available in the market cannot be fractured. In an ex vivo test, the possibility of ring fracture of aortic valve bioprostheses produced in Brazil when submitted to radial force application using a high-pressure non-compliant balloon was evaluated.

Methods: One unit of each aortic valve bioprosthesis model, sizes 19 and 21 mm, produced by Brazilian companies (Braile Biomédica, Cardioprótese, and Labcor), was used. In the experiment, a non-compliant high-pressure balloon (Atlas®-Gold), 1 mm larger than the external diameter of the prosthesis, was positioned inside the valve annulus and inflated gradually aiming to fracture the prosthesis. Fracture pressures and photographic and radiological images of the prostheses before and after test were recorded.

Results: All prostheses were fractured. In the models with metal ring, the fracture pressures were between 23 and 25 atm. In the other prostheses, the rupture occurred between 10 and 13 atm. No deformations in the structure were observed, which could potentially damage the aortic root.

Conclusion: All the Brazilian prostheses evaluated were fractured, although the presence of a metal ring in the prosthesis framework increases the pressure required for fracture. The information obtained helps in the planning of valve-in-valve procedures in patients with aortic valve bioprostheses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9162417PMC
http://dx.doi.org/10.21470/1678-9741-2021-0272DOI Listing

Publication Analysis

Top Keywords

aortic valve
20
ring fracture
16
valve bioprostheses
16
non-compliant high-pressure
8
vivo test
8
valve-in-valve procedures
8
fracture pressures
8
metal ring
8
aortic
6
valve
6

Similar Publications

Reversed valved Potts shunt for refractory primary pulmonary arterial hypertension.

Multimed Man Cardiothorac Surg

January 2025

• Pediatric and Congenital Cardiac Surgery, LMU University Hospital, Munich, Germany • Congenital Cardiac Surgery, German Heart Center Munich, Munich, Germany • European Pediatric Heart Center EKHZ Munich, Munich, Germany.

This procedure is carried out via a full sternotomy using standard aortic and bicaval cannulations. For the aortic and pulmonary anastomoses, selective antegrade unilateral cerebral perfusion is used after cooling the body temperature to 26 °Celsius. A 12-mm Hancock conduit is interposed between the pulmonary artery and the proximal descending aorta using standard running suture techniques.

View Article and Find Full Text PDF

Background: The coexistence of transthyretin cardiac amyloidosis (ATTR-CA) and aortic stenosis (AS) is increasingly recognized, but the clinical consequences are unclear. We aimed to characterize clinical outcomes in AS plus ATTR-CA compared with only AS or ATTR-CA.

Methods And Results: In a retrospective cohort study, patients with AS only, ATTR-CA only, or AS plus ATTR-CA were identified using all-payer claims data (2015-2021).

View Article and Find Full Text PDF

Valvular Heart Disease Care in Pakistan: Impact of the Multidisciplinary Valve Heart Team.

JACC Adv

December 2024

Department of Interventional Cardiology& structural heart interventions, NICVD, Karachi, Pakistan.

Background: Patients with complex valvular heart disease (VHD) should be evaluated by a multidisciplinary heart team (HT). In low- and middle-income countries, referral practices are more variable, permitting any physician to refer patients directly to a cardiac surgeon without prior formal evaluation by a cardiologist with expertise in VHD.

Objectives: The goal of the study was to examine the demographics of VHD patients seen in a large heart valve center in a low- and middle-income country and to assess the impact of the multidisciplinary HT in patients referred for valve surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!