Purpose The aim of this study was to evaluate the accuracy of bladder pressures in the diagnosis and management of abdominal compartment syndrome (ACS). Methods After Institutional Review Board (IRB) approval, nine operative abdominal trauma patients were prospectively studied over an 18-month period. Bladder pressures were compared to pressures obtained from intra-operatively placed electronic transducer located in the pelvis. Statistical analysis was performed using methods described by Bland and Altman. Results A Bland-Altman plot was used to assess the differences between bladder and transducer pressures. There was little agreement between the two methods at low (10-15 mmHg) and high (30-70 mmHg) pressures. At higher pressures, there was a notable difference between these two types of pressure. No patients required repeated operation. One patient died from severe traumatic brain injury. Conclusion Measurements obtained from the urinary bladder did not agree well with those obtained from within the peritoneal cavity. Furthermore, abdominal pressures greater than 20 mmHg did not show signs of ACS in this selected population, raising doubts about the utility of using abdominal pressures alone in the management of ACS.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9116434 | PMC |
http://dx.doi.org/10.7759/cureus.24238 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!