Analytical performances of a glycated albumin assay that is traceable to standard reference materials and reference range determination.

J Clin Lab Anal

Diagnostics Department, Asahi Kasei Pharma Corporation, IVD Kit Product Group, Yurakucho, Japan.

Published: July 2022

Background: Glycated albumin (GA) is an intermediate-term marker for monitoring glycemic control (preceding 2-3 weeks) in patients with diabetes mellitus. We evaluated the performance of Lucica Glycated Albumin-L, a new GA assay that is traceable to standard reference materials and determined the reference range in healthy subjects without diabetes.

Methods: The performance and reference range studies were conducted in accordance with Clinical and Laboratory Standards Institute (CLSI) Guidelines. The traceability was established using reference material recommended by the Japan Society of Clinical Chemistry (JSCC).

Results: The coefficient of variation (CV) of overall repeatability, within-laboratory precision, and overall reproducibility values of GA values were not more than 2.6%, 3.3%, and 1.6%, respectively, among laboratories. The GA values showed good linearity from 173 to 979 mmol/mol (9.4%-54.9%) across the assay range. The GA reference range in 262 healthy subjects was between 183 and 259 mmol/mol (9.9%-14.2%) while that of subjects with diabetes was 217-585 mmol/mol (11.8-32.6%). The reagent was stable for 2 months on the bench at room temperature. The limits of blank, detection, and qualification were 6.9, 7.9, and 9.7 μmol/L for GA concentration, and 3.8, 7.0, and 21.8 μmol/L for albumin concentration, respectively. Hemoglobin slightly affected the assay, while other classical interfering substances had no significant impact.

Conclusions: The present GA assay shows comparable performance to current clinical assays and could be used for intermediate-term monitoring of glycemic control in diabetes patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9280011PMC
http://dx.doi.org/10.1002/jcla.24509DOI Listing

Publication Analysis

Top Keywords

reference range
16
glycated albumin
8
assay traceable
8
traceable standard
8
standard reference
8
reference materials
8
monitoring glycemic
8
glycemic control
8
healthy subjects
8
reference
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!