A wireless ultrasound patch detects mild-to-moderate central hypovolemia during lower body negative pressure.

J Trauma Acute Care Surg

From the Health Sciences North Research Institute (J.-É.S.K., A.M.E., J.K.E.); Northern Ontario School of Medicine (J.K.E.), Sudbury, Ontario, Canada; Flosonics Medical (J.-É.S.K., M.E, Z.Y., A.M.E., J.K.E.), Toronto, Ontario, Canada; and Human Integrative and Environmental Physiology Laboratory, Department of Cardiovascular Diseases (C.-H.K. B.D.J.), Mayo Clinic, Rochester, MN.

Published: August 2022

Background: We have developed a wireless, wearable Doppler ultrasound system that continuously measures the common carotid artery Doppler pulse. A novel measure from this device, the Doppler shock index, accurately detected moderate-to-severe central blood volume loss in a human hemorrhage model generated by lower body negative pressure. In this analysis, we tested whether the wearable Doppler could identify only mild-to-moderate central blood volume loss.

Methods: Eleven healthy volunteers were recruited and studied in a physiology laboratory at the Mayo Clinic. Each participant underwent a lower body negative protocol in duplicate. Carotid Doppler measures including Doppler shock indices were compared with blood pressure and the shock index for their ability to detect both 10% and 20% reductions in stroke volume.

Results: All carotid Doppler measures were better able to detect diminishing stroke volume than either systolic or mean arterial pressure. Falling carotid artery corrected flow time and rising heart rate/corrected flow time (DSI FTc ) were the most sensitive measures for detecting 10% and 20% stroke volume reductions, respectively. The area under the receiver operator curves (AUROCs) for all shock indices was at least 0.86; however, the denominators of the two Doppler shock indices (i.e., the corrected flow time and velocity time integral) had AUROCs ranging between 0.81 and 0.9, while the denominator of the traditional shock index (i.e., systolic blood pressure) had AUROCs between 0.54 and 0.7.

Conclusion: The wearable Doppler ultrasound was able to continuously measure the common carotid artery Doppler pulse. Carotid Doppler measures were highly sensitive at detecting both 10% and 20% stroke volume reduction. All shock indices performed well in their diagnostic ability to measure mild-to-moderate central volume loss, although the denominators of both Doppler shock indices individually outperformed the denominator of the traditional shock index.

Level Of Evidence: Diagnostic test or criteria; Level III.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9323554PMC
http://dx.doi.org/10.1097/TA.0000000000003698DOI Listing

Publication Analysis

Top Keywords

shock indices
20
doppler shock
16
mild-to-moderate central
12
lower body
12
body negative
12
doppler
12
wearable doppler
12
carotid artery
12
carotid doppler
12
doppler measures
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!