The Mechanical Behavior of a Screwless Morse Taper Implant-Abutment Connection: An In Vitro Study.

Materials (Basel)

Medical Center-University of Freiburg, Center for Dental Medicine, Department of Prosthetic Dentistry, Faculty of Medicine, University of Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany.

Published: May 2022

The use of screwless Morse taper implant−abutment connections (IAC) might facilitate the clinician’s work by eliminating the mechanical complications associated with the retention screw. The aim of this study is to evaluate the effect of artificial chewing on the long-term stability of screwless Morse taper IACs. Thirty-two implant abutments restored with an upper central incisor zirconia crown were used and divided into four groups according to the implant−abutment assembling manner (C1,H: screw retained (20 Ncm); C2: tapped; or C3: torqued (20 Ncm; the screws were removed before the dynamic loading)). All specimens were subjected to a cyclic loading (98 N) for 10 million chewing cycles. The survived samples were exposed to a pull-off force until failure/disassembling of the connection. All the samples revealed a 100% survival. Regarding the pull-off test, the screw-retained internal hexagonal IAC revealed significantly higher resistance to failure/disassembling (769.6 N) than screwless conical IACs (171.6 N−246 N) (p < 0.0001). The retention forces in the Morse taper groups were not significantly different (p > 0.05). The screw-retained hexagonal IAC showed the highest retention stability. The screw preload/retention in the conical IAC was lost over time in the group where the screws were kept in place during loading. Nevertheless, the screwless Morse taper IACs were stable for an extended service time and might represent a valid form of treatment for single-tooth replacement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9103680PMC
http://dx.doi.org/10.3390/ma15093381DOI Listing

Publication Analysis

Top Keywords

morse taper
20
screwless morse
16
taper iacs
8
hexagonal iac
8
screwless
5
morse
5
taper
5
mechanical behavior
4
behavior screwless
4
taper implant-abutment
4

Similar Publications

Total hip arthroplasty (THA) is a very successful operation. Once the problem of implant fixation was solved with the use of bone cement, the next development steps focused on improving the bearing. Weber, a Swiss surgeon, introduced the first modular heads in THA.

View Article and Find Full Text PDF

Purpose: This study assessed the microgap width and adhesion of three bacterial species in four dental implants with different interlocks under four screwing torques.

Materials And Methods: Ten samples of four implant systems with various interlockings, including full-hexagonal (FHI), cylindrical-conical trilobe-index (TLI), Morse-taper with octagon terminal index (OI), and hexagonal interlock (slip-fit) (HI-SF), were used. The abutments were screwed to the fixtures under torques of 10, 20, 30, and 40 Ncm.

View Article and Find Full Text PDF

The purpose of this study was to evaluate the effect of a single dose of ionizing radiation (30 Gy) on the osseointegration of implants in the rabbit tibia. Twenty rabbits received two dental Morse-tapered junction implants and one implant in each tibia. The animals were randomly divided into two groups (n=10), non-irradiated (NoIr) and irradiated (Ir), wherein the Ir group received a single dose of 30 Gy radiation 2 weeks after implant installation.

View Article and Find Full Text PDF

Statement Of Problem: The biomechanical stability of the implant-prosthesis assembly and its maintenance under function is a determining factor in the success of implant prosthesis rehabilitation, but studies of different angular tapers are lacking.

Purpose: The purpose of this in vitro study was to evaluate the axial displacement and torque loss of prosthetic abutments with Morse-type connections of different angular tapers after thermomechanical cycling by using microcomputed tomography (µCT) and a digital torque wrench.

Material And Methods: Eighteen Ø3.

View Article and Find Full Text PDF

The present study evaluated the mechanical behavior of five designs of Morse taper (MT) connections with and without the application of loads. For this, the detorque of the fixing screw and the traction force required to disconnect the abutment from the implant were assessed. A total of 100 sets of implants/abutments (IAs) with MT-type connections were used, comprising five groups ( = 20/group): (1) Group Imp 11.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!