Comparison of Doppler Flow Velocity and Thermodilution Derived Indexes of Coronary Physiology.

JACC Cardiovasc Interv

British Heart Foundation Centre of Excellence and National Institute for Health Research Biomedical Research Centre at the School of Cardiovascular Medicine and Sciences, King's College London, London, United Kingdom. Electronic address:

Published: May 2022

Objectives: The aim of this study was to compare Doppler flow velocity and thermodilution-derived indexes and to determine the optimal thermodilution-based diagnostic thresholds for coronary flow reserve (CFR).

Background: The majority of clinical data and diagnostic thresholds for flow-based indexes are derived from Doppler measurements, and correspondence with thermodilution-derived indices remain unclear.

Methods: An international multicenter registry was conducted among patients who had coronary flow measurements using both Doppler and thermodilution techniques in the same vessel and during the same procedure.

Results: Physiological data from 250 vessels (in 149 patients) were included in the study. A modest correlation was found between thermodilution-derived CFR (CFR) and Doppler-derived CFR (CFR) (r = 0.36; P < 0.0001). CFR overestimated CFR (mean 2.59 ± 1.46 vs 2.05 ± 0.89; P < 0.0001; mean bias 0.59 ± 1.24 by Bland-Altman analysis), the relationship being described by the equation CFR = 1.04 × CFR + 0.50. The commonly used dichotomous CFR threshold of 2.0 had poor sensitivity at predicting a CFR value <2.5. The optimal CFR threshold was 2.5 (sensitivity 75.54%, specificity 81.25%). There was only a weak correlation between hyperemic microvascular resistance and index of microvascular resistance (r = 0.19; P < 0.0001), due largely to variation in the measurement of flow by each modality. Forty-four percent of patients were discordantly classified as having abnormal microvascular resistance by hyperemic microvascular resistance (≥2.5 mm Hg · cm · s) and index of microvascular resistance (≥25).

Conclusions: CFR calculated by thermodilution overestimates Doppler-derived CFR, while both parameters show modest correlation. The commonly used CFR threshold of 2.0 has poor sensitivity for identifying vessels with diminished CFR, but using the same binary diagnostic threshold as for Doppler (<2.5) yields reasonable diagnostic accuracy. There was only a weak correlation between microvascular resistance indexes assessed by the 2 modalities.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9126183PMC
http://dx.doi.org/10.1016/j.jcin.2022.03.015DOI Listing

Publication Analysis

Top Keywords

doppler flow
8
flow velocity
8
diagnostic thresholds
8
coronary flow
8
cfr
8
cfr cfr
8
p < 00001
8
comparison doppler
4
flow
4
velocity thermodilution
4

Similar Publications

Pelvic venous disorder (PVD) is a common gynecological disorder caused by chronic pelvic venous congestion. Clinically, it primarily manifests as lower abdominal pain, lower back pain, dyspareunia, increased menstrual flow, and increased vaginal discharge, though positive physical signs are few. The gynecological examination may reveal bilateral ovarian tenderness.

View Article and Find Full Text PDF

Microplastics (MP) are known to be ubiquitous. The pathways and fate of these contaminants in the marine environment are receiving increasing attention, but still knowledge gaps exist. In particular, the link between mass-based MP quantification and oceanographic parameters is often lacking.

View Article and Find Full Text PDF

To what extent sildenafil, a selective inhibitor of the type-5 phosphodiesterase modulates systemic redox status and cerebrovascular function during acute exposure to hypoxia remains unknown. To address this, 12 healthy males (aged 24 ± 3 y) participated in a randomized, placebo-controlled crossover study involving exposure to both normoxia and acute (60 min) hypoxia (Fi = 0.14), followed by oral administration of 50 mg sildenafil and placebo (double-blinded).

View Article and Find Full Text PDF

Retrospective Analysis of the Correlation between Umbilical Blood Flow Index and Maternal and Fetal Outcomes in Pregnant Women with Gestational Diabetes.

Br J Hosp Med (Lond)

January 2025

Department of Obstetrics and Gynecology, The First Clinical Medical College of Three Gorges University, Yichang Central People's Hospital, Yichang, Hubei, China.

Gestational diabetes mellitus (GDM) is a common complication during pregnancy. This retrospective study investigates the correlation between umbilical blood flow index and maternal-fetal outcomes in pregnant women with GDM, aiming to contribute to evidence-based risk assessment and management strategy in this high-risk obstetric population. This retrospective study recruited 119 pregnant women with GDM who were admitted to the Yichang Central People's Hospital, between January 2022 and January 2024.

View Article and Find Full Text PDF

The autologous reconstruction of the female breast using a microsurgical DIEP flap is a reliable and safe method. To detect impairments early and preserve the microvascular flap through timely revision, a better understanding of physiologic perfusion dynamics is necessary. This exploratory study examines changes in microcirculation in free DIEP flaps within the first 72 h after vascular anastomosis using laser Doppler flowmetry and white-light spectrophotometry.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!