The aim of the study was to validate the performance of the Optomed Aurora handheld fundus camera in diabetic retinopathy (DR) screening. Patients who were affected by diabetes mellitus and referred to the local DR screening service underwent fundus photography using a standard table-top fundus camera and the Optomed Aurora handheld fundus camera. All photos were taken by a single, previously unexperienced operator. Among 423 enrolled eyes, we found a prevalence of 3.55% and 3.31% referable cases with the Aurora and with the standard table-top fundus camera, respectively. The Aurora obtained a sensitivity of 96.9% and a specificity of 94.8% in recognizing the presence of any degree of DR, a sensitivity of 100% and a specificity of 99.8% for any degree of diabetic maculopathy (DM) and a sensitivity of 100% and specificity of 99.8% for referable cases. The overall concordance coefficient k (95% CI) was 0.889 (0.828-0.949) and 0.831 (0.658-1.004) with linear weighting for DR and DM, respectively. The presence of hypertensive retinopathy (HR) was recognized by the Aurora with a sensitivity and specificity of 100%. The Optomed Aurora handheld fundus camera proved to be effective in recognizing referable cases in a real-life DR screening setting. It showed comparable results to a standard table-top fundus camera in DR, DM and HR detection and grading. The Aurora can be integrated into telemedicine solutions and artificial intelligence services which, in addition to its portability and ease of use, make it particularly suitable for DR screening.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9103652PMC
http://dx.doi.org/10.3390/jcm11092352DOI Listing

Publication Analysis

Top Keywords

fundus camera
32
handheld fundus
16
table-top fundus
16
optomed aurora
12
aurora handheld
12
standard table-top
12
referable cases
12
camera
8
camera diabetic
8
diabetic retinopathy
8

Similar Publications

Objectives: Fundoscopy is crucial in the emergency department to identify or rule out serious ocular and neurological conditions. Despite its clinical importance, fundoscopy is often omitted due to the technical challenges associated with traditional direct ophthalmoscopy, particularly for non-ophthalmologists. This study examines emergency physicians' practices, confidence levels, and training related to various modalities of fundoscopy including traditional direct ophthalmoscopes, binocular indirect ophthalmoscopes, panoptic ophthalmoscopes, slit lamp fundoscopy and fundus cameras; and explores the potential role of alternative modalities, such as fundus cameras, in Canadian emergency departments.

View Article and Find Full Text PDF

The aim of the study was to determine the thickness of choroidal layers in mixed breed dogs suffering from retinal atrophy (RA) and showing symptoms of progressive retinal atrophy (PRA), with the use of SD-OCT. The study was performed on 50 dogs divided into two groups: 25 dogs diagnosed with retinal atrophy (RA) with PRA symptoms aged 1.5-14 years and 25 healthy dogs aged 2-12 years.

View Article and Find Full Text PDF

Where adopted, Autonomous artificial Intelligence (AI) for Diabetic Retinal Disease (DRD) resolves longstanding racial, ethnic, and socioeconomic disparities, but AI adoption bias persists. This preregistered trial determined sensitivity and specificity of a previously FDA authorized AI, improved to compensate for lower contrast and smaller imaged area of a widely adopted, lower cost, handheld fundus camera (RetinaVue700, Baxter Healthcare, Deerfield, IL) to identify DRD in participants with diabetes without known DRD, in primary care. In 626 participants (1252 eyes) 50.

View Article and Find Full Text PDF

Background: Ocular proton beam therapy (OPT) planning would benefit from an accurate incorporation of fundus photographs, as various intra-ocular structures, such as the fovea, are not visible on conventional modalities such as Magnetic Resonance Imaging (MRI). However, the use of fundus photographs in OPT is limited, as the eye's optics induce a nonuniform patient-specific deformation to the images.

Purpose: To develop a method to accurately map fundus photographs to three-dimensional images.

View Article and Find Full Text PDF
Article Synopsis
  • * Researchers analyzed medical records of 100 eyes with AH, finding that 76 had other retinal diseases; TC-cSO showed the best results, with 94% of images having minimal obscuration from asteroid bodies.
  • * TC-cSO outperformed the other devices in accuracy (95%) and sensitivity (81%) for diagnosing retinal conditions, making it an effective alternative for screening when
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!