Purpose: This study aimed to evaluate the shielding integrity of a typical radiotherapy facility using the Monte Carlo (MC) method.

Materials And Methods: EGSnrc MC code was used to design a radiotherapy bunker with appropriate materials and thicknesses. A concrete density of 2.36 g/cm was used as a shielding material for primary and secondary barriers. The lead slab was used in the entrance door. The complex geometries of the bunker were modeled by using the egs++ application code embedded in the software. Phase-space generated from a linac machine built with BEAMnrc codes was used as a source of 18 MV X-ray beam set at 100 cm source-surface distance with a field size of 40 cm × 40 cm. Energy deposited in each geometrical region was evaluated and analyzed.

Results: Energy deposited at the entrance door, supervised and controlled areas were found to be approximately 0%. No significant difference in the energy deposition on the geometries was observed when the gantry angles were set at either 90° or 270° ( = 1).

Conclusion: The findings in this study revealed that the EGSnrc MC code can be used as a veritable tool in the design and evaluation of structural shielding efficiency in a radiotherapy facility.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9084580PMC
http://dx.doi.org/10.4103/jmp.jmp_76_21DOI Listing

Publication Analysis

Top Keywords

radiotherapy facility
12
design evaluation
8
evaluation structural
8
structural shielding
8
typical radiotherapy
8
monte carlo
8
egsnrc code
8
entrance door
8
energy deposited
8
shielding
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!