The relationship between age of digital mammography systems and number of reported faults and downtime.

Phys Med

National Coordinating Centre for the Physics in Mammography (NCCPM), Level B, St Luke's Wing, Royal Surrey County Hospital, Guildford GU2 7XX, UK.

Published: June 2022

Purpose: To investigate the relationship between age of mammographic x-ray equipment, and number of reported faults and related consequences.

Methods: A centralised online fault reporting database is used by all UK breast screening programmes to collate faults with mammography equipment. Data on faults occurring in 2018 and 2019 for digital x-ray imaging systems were analysed. The effect of the age of mammography systems on the number of equipment faults, and the consequences of these faults was examined. The number of days downtime, number of cancelled appointments, number of repeated images, and number of recalled participants were used to quantify the severity of faults.

Results: This analysis covers a two year period and includes 4271 faults and 522 individual x-ray sets. On average, an x-ray set was 6.1 years old at the time when a fault occurred. 77% of x-ray sets experienced five of fewer annual faults. X-ray sets of nine years old had the highest average number of annual faults. Systems of ten years old had the highest average number of days downtime per year, and the highest average number of cancellations per year. The indicated primary use of 48% of the x-ray sets included in this analysis was screening, but a disproportionate 87% of cancelled appointments occurred due to faults on these units compared to those used primarily for assessment, or for a mixture of assessment and screening.

Conclusions: Information from this unique dataset can be used to support guidance on equipment replacement programmes for mammographic x-ray sets.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2022.04.015DOI Listing

Publication Analysis

Top Keywords

x-ray sets
20
highest average
12
average number
12
number
10
faults
10
relationship age
8
mammography systems
8
systems number
8
number reported
8
reported faults
8

Similar Publications

Objective: This investigation attempted to examine the effectiveness of CT-derived peritumoral and intratumoral radiomics in forecasting microsatellite instability (MSI) status preoperatively among gastric cancer (GC) patients.

Methods: A retrospective analysis was performed on GC patients from February 2019 to December 2023 across three healthcare institutions. 364 patients (including 41 microsatellite instability-high (MSI-H) and 323 microsatellite instability-low/stable (MSI-L/S)) were stratified into a training set (n = 202), an internal validation set (n = 84), and an external validation set (n = 78).

View Article and Find Full Text PDF

Purpose: To construct a nomogram combining Kaiser score (KS), synthetic MRI (syMRI) parameters, apparent diffusion coefficient (ADC), and clinical features to distinguish benign and malignant breast lesions better.

Methods: From December 2022 to February 2024, a retrospective cohort of 168 patients with breast lesions diagnosed as Breast Imaging Reporting and Data System (BI-RADS) category 4 by ultrasound and/or mammography was included. The research population was divided into the training set (n = 117) and the validation set (n = 51) by random sampling with a ratio of 7:3.

View Article and Find Full Text PDF

Introduction: A large number of middle-aged and elderly patients have an insufficient understanding of osteoporosis and its harm. This study aimed to establish and validate a convolutional neural network (CNN) model based on unenhanced chest computed tomography (CT) images of the vertebral body and skeletal muscle for opportunistic screening in patients with osteoporosis.

Materials And Methods: Our team retrospectively collected clinical information from participants who underwent unenhanced chest CT and dual-energy X-ray absorptiometry (DXA) examinations between January 1, 2022, and December 31, 2022, at four hospitals.

View Article and Find Full Text PDF

Early-stage cutaneous melanoma patients generally have a favorable prognosis, yet a significant proportion of metastatic melanoma cases arise from this group, highlighting the need for improved risk stratification using novel prognostic biomarkers. The Dutch Early-Stage Melanoma (D-ESMEL) study introduces a robust, population-based methodology to develop an absolute risk prediction model for stage I/II melanoma, incorporating clinical, imaging, and multi-omics data to identify patients at increased risk for distant metastases. Utilizing the Netherlands Cancer Registry and Dutch Nationwide Pathology Databank, we collected primary tumor samples from early-stage melanoma patients, with and without distant metastases during follow-up.

View Article and Find Full Text PDF

Trait mindfulness refers to one's disposition or tendency to pay attention to their experiences in the present moment, in a non-judgmental and accepting way. Trait mindfulness has been robustly associated with positive mental health outcomes, but its neural underpinnings are poorly understood. Prior resting-state fMRI studies have associated trait mindfulness with within- and between-network connectivity of the default-mode (DMN), fronto-parietal (FPN), and salience networks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!