Study Objectives: The most sensitive and specific investigative method for the diagnosis of narcolepsy type 1 (NT1) is the determination of hypocretin-1 (orexin-A) deficiency (≤110 pg/mL) in cerebrospinal fluid using a radioimmunoassay (RIA). We aimed to assess the reliability of the Phoenix Pharmaceuticals hypocretin-1 RIA, by determining the lower limit of quantification (LLOQ), the variability around the cutoff of 110 pg/mL, and the inter- and intra-assay variability.

Methods: Raw data of 80 consecutive hypocretin-1 RIAs were used to estimate the intra- and inter-assay coefficient of variation (CV). The LLOQ was established and defined as the lowest converted concentration with a CV <25%; the conversion is performed using a harmonization sample which is internationally used to minimize variation between RIAs.

Results: The mean intra-assay CV was 4.7%, while the unconverted inter-assay CV was 28.3% (18.5% excluding 2 outliers) and 7.5% when converted to international values. The LLOQ was determined as 27.9 pg/mL. The intra-assay CV of RIAs with lower specific radioactive activity showed a median of 5.6% (n = 41, range 1.6%-17.0%), which was significantly higher than in RIAs with higher specific activity (n = 36; median 3.2%, range 0.4%-11.6%, p = .013). The CV around the 110 pg/mL cutoff was <7%.

Conclusions: Hypocretin-1 RIAs should always be harmonized using standard reference material. The specific activity of an RIA has a significant impact on its reliability, because of the decay of 125I radioactivity. Values around the hypocretin-1 cut-off can reliably be measured. Hypocretin-1 concentrations below 28 pg/mL should be reported as "undetectable" when measured with the Phoenix Pharmaceuticals RIA.

Clinical Trial Information: This study is not registered in a clinical trial register, as it has a retrospective database design.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9272241PMC
http://dx.doi.org/10.1093/sleep/zsac103DOI Listing

Publication Analysis

Top Keywords

cerebrospinal fluid
8
fluid radioimmunoassay
8
limit quantification
8
hypocretin-1
4
hypocretin-1 measurements
4
measurements cerebrospinal
4
radioimmunoassay assay
4
assay reliability
4
reliability limit
4
quantification study
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!