Context: Continuous glucose monitoring (CGM) is increasingly being used both for day-to-day management in patients with diabetes and in clinical research. While data on glycemic profiles of healthy, nondiabetic individuals exist, data on nondiabetic very young children are lacking.

Objective: This work aimed to establish reference sensor glucose ranges in healthy, nondiabetic young children, using a current-generation CGM sensor.

Methods: This prospective observational study took place in an institutional practice with healthy, nondiabetic children aged 1 to 6 years with normal body mass index. A blinded Dexcom G6 Pro CGM was worn for approximately 10 days by each participant. Main outcome measures included CGM metrics of mean glucose, hyperglycemia, hypoglycemia, and glycemic variability.

Results: Thirty-nine participants were included in the analyses. Mean average glucose was 103 mg/dL (5.7 mmol/L). Median percentage time between 70 and 140 mg/dL (3.9-7.8 mmol/L) was 96% (interquartile range, 92%-97%), mean within-individual coefficient of variation was 17 ± 3%, median time spent with glucose levels greater than 140 mg/dL was 3.4% (49 min/day), and median time less than 70 mg/dL (3.9 mmol/L) was 0.4% (6 min/day).

Conclusion: Collecting normative sensor glucose data and describing glycemic measures for young children fill an important informational gap and will be useful as a benchmark for future clinical studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9049110PMC
http://dx.doi.org/10.1210/jendso/bvac060DOI Listing

Publication Analysis

Top Keywords

healthy nondiabetic
16
young children
16
nondiabetic young
12
continuous glucose
8
glucose monitoring
8
profiles healthy
8
sensor glucose
8
mg/dl mmol/l
8
140 mg/dl
8
median time
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!