Rational And Objectives: Non-invasive cardiac output (CO) measurements are essential during the immediate post-operative course of young, congenital heart repaired patients. The use of the Ultrasonic Cardiac Output Monitor (USCOM) in pediatric intensive care units (PICU) is increasing. The literature on accuracy of USCOM in young, critically ill, mechanically ventilated, hemodynamically supported patients is scarce. We aimed to assess agreement between the USCOM device and echocardiography for measurements of CO in this population.

Materials (patients) And Methods: A prospective observational study in a pediatric cardiac intensive care unit (PCICU). Paired CO measurements were taken in young, mechanically ventilated, immediate post-operative patients with exclusion of unrepaired or residual intra-cardiac shunt, using USCOM and echocardiography, by two separate senior performers. Agreement between echocardiography and USCOM was assessed by percentage error and Bland-Altman analysis.

Results: One hundred and thirteen comparison scans were performed on 61 patients: mean age 94 ± 111 d, weight 4.7 ± 2.1 kg, vaso-inotropic score 15.3 ± 11, and STAT score 3-4 (46%). Mean USCOM cardiac index (CI) percent difference was -9.6% (45.6) and velocity-time-integral (VTI) 8.9% (34.7). Bland-Altman analyzes demonstrated poor agreement comparing USCOM to echocardiography with regard to CI, stroke volume (SV), VTI and aortic diameter (AO) measurements.

Conclusion: Our study shows that USCOM underestimates CI in comparison with echocardiography; therefore USCOM should be used with great caution as an absolute estimate or surrogate of CI in neonates and infants in the immediate post-operative, congenital heart surgery period.

Download full-text PDF

Source
http://dx.doi.org/10.1177/08850666221099830DOI Listing

Publication Analysis

Top Keywords

uscom
9
cardiac output
8
congenital heart
8
intensive care
8
mechanically ventilated
8
uscom echocardiography
8
echocardiography uscom
8
echocardiography
6
patients
6
cardiac
5

Similar Publications

Multiple sclerosis (MS) is a chronic neurological condition resulting in decreased aerobic capacity (peak VO). The hemodynamic responses to peak exercise in MS are unknown. Further, it is unknown if the hemodynamic responses are due to disease or fitness.

View Article and Find Full Text PDF

To (1) assess the accuracy of the Uscom BP+ oscillometric upper-arm professional blood pressure (BP) monitor in children and adolescents as per the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018); (2) assess the suitability of a non-standard cuff; and (3) assess the impact of motion artefact on device accuracy and precision. Children and adolescents were recruited to fulfil the sex/cuff distribution criteria of the Universal Standard. Single arm sequential BP measurements were used, the test device measured on deflate with an altered child/adolescent algorithm, using three standard (Extra Small, Small Adult, Adult) and, as exploratory analysis, additionally in one non-standard cuff (Wide Range; WR) (arm circumferences: 12-17, 16-24, 22-32, 22-42 cm, respectively).

View Article and Find Full Text PDF

Introduction: Targeted therapy has improved clinical outcomes for various types of cancer. However, their use is associated with dermatologic adverse events that impact quality of life and consistent therapies.

Objectives: The US Cutaneous Oncodermatology Management (USCOM) multidisciplinary-guided algorithm for preventing and managing cutaneous targeted therapy-related adverse events provides practical recommendations for cancer patients and survivors.

View Article and Find Full Text PDF

Proteome-Wide Genetic Investigation of Large Artery Stiffness.

JACC Basic Transl Sci

October 2024

Division of Cardiovascular Medicine, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania, USA.

The molecular mechanisms contributing to large artery stiffness (LAS) are not fully understood. The aim of this study was to investigate the association between circulating plasma proteins and LAS using complementary proteomic and genomic analyses. A total of 106 proteins associated with carotid-femoral pulse-wave velocity, a noninvasive measure of LAS, were identified in 1,178 individuals from the Asklepios study cohort.

View Article and Find Full Text PDF

Background: Women developing preeclampsia (PE) or fetal growth restriction (FGR) during pregnancy are at higher risk for cardiovascular diseases (CVD) later in life. We aimed to analyse cardiovascular health of women 10-20 years after affected pregnancies in comparison to women after uneventful pregnancies. In addition, we assessed a potential long-term effect of the NO-donor pentaerythrityl tetranitrate (PETN).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!