Background: Measurements of liver stiffness obtained with 2-dimensional shear wave elastography (2D-SWE) have been widely used to clinically assess liver fibrosis. However, differences between different 2D-SWE systems can lead to confusion when interpreting measurements. This study investigated the variability between a recently released sound touch elastography (STE) system and a supersonic shear imaging (SSI) system and assessed the degree of intersystem discrepancy using the different liver stiffness value (LSV) thresholds recommended by the Society of Radiologists in Ultrasound (SRU) for assessing liver fibrosis.
Methods: A total of 4,152 patients who had undergone STE and SSI on the same day were enrolled in this retrospective study. First, intrasystem agreement for STE and SSI was assessed. Then, intraclass correlation coefficients (ICCs) and the Bland-Altman method were used to assess intersystem variability for all cases, classified according to the thresholds recommended by the SRU. The effects of age, gender, and body-mass index (BMI) were evaluated using multivariate linear regression analysis and attributive intervals were computed for STE and SSI at each of the different thresholds.
Results: The ICCs for STE and SSI intrasystem agreement were 0.94 [95% confidence interval (CI): 0.937-0.943; P<0.001] and 0.984 (95% CI: 0.984-0.985; P<0.001), respectively. The 95% limit of agreement (LOA) for all cases ranged from -6.96 to 7.44 kPa. The 95% LOA increased as the threshold values rose, and intersystem variability was obvious, even at the smallest threshold (the 95% LOA at values ≤5 kPa was -0.85 to 2.08 kPa, while that at values >17 kPa was -20.81 to 14.71 kPa). The adjusted R for age, gender, and BMI was only 0.018 (all P value <0.05).
Conclusions: There was clear variability between STE and SSI, in contrast with some previous studies with small sample sizes, and consistent with others. Intersystem variability increased with the elevation of the LSV thresholds recommended by the SRU. Gender and BMI had little effect on intersystem variability. Future research could compare STE and SSI in different liver diseases, assessing the feasibility of the SRU-recommended thresholds in proven pathologies and evaluating the test-retest repeatability.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9014163 | PMC |
http://dx.doi.org/10.21037/qims-21-863 | DOI Listing |
Injury
August 2023
Wright State University Department of Orthopaedic Surgery, 30 E. Apple St., Ste 2200, Dayton, Ohio 45409 United States of America.
Introduction: The prophylactic intravenous antibiotic regimen for Gustilo-Anderson Type III open fractures traditionally consists of cefazolin with an aminoglycoside plus penicillin for gross contamination. Cefotetan, a second-generation cephalosporin, offers a wide spectrum of activity against both aerobes and anaerobes as well as against Gram-positive and Gram-negative bacteria. Cefotetan has not been previously established within orthopedic surgery as a prophylactic intravenous agent.
View Article and Find Full Text PDFJ Gastrointest Surg
October 2022
Department of Surgery, McGill University, Montreal, QC, Canada.
Background: Preoperative administration of oral antibiotic bowel preparation (OABP) alone has been shown to reduce infectious outcomes in patients undergoing elective colectomy. However, it remains unclear if these benefits extend to the emergency setting. This is a retrospective, propensity-score matched study comparing 30-day perioperative morbidity between those who received OABP alone versus no preparation prior to urgent colectomy.
View Article and Find Full Text PDFJ Plast Reconstr Aesthet Surg
September 2022
Division of Plastic and Reconstructive Surgery, Department of Surgery, Stanford University Medical Center, Stanford, CA, United States. Electronic address:
Background: Textured tissue expanders (TTEs) were introduced to limit migration and reduce capsular contracture, which were inherent to smooth tissue expanders (STEs). Previous reports suggest that textured devices have increased rates of bacterial contamination and biofilm formation in comparison with smooth devices. Recently, the relative increased association of anaplastic large cell lymphoma (ALCL) with textured versus smooth devices has led to increased adoption of smooth devices.
View Article and Find Full Text PDFQuant Imaging Med Surg
May 2022
Department of Ultrasound, West China Hospital of Sichuan University, Chengdu, China.
Background: Measurements of liver stiffness obtained with 2-dimensional shear wave elastography (2D-SWE) have been widely used to clinically assess liver fibrosis. However, differences between different 2D-SWE systems can lead to confusion when interpreting measurements. This study investigated the variability between a recently released sound touch elastography (STE) system and a supersonic shear imaging (SSI) system and assessed the degree of intersystem discrepancy using the different liver stiffness value (LSV) thresholds recommended by the Society of Radiologists in Ultrasound (SRU) for assessing liver fibrosis.
View Article and Find Full Text PDFFront Psychol
December 2021
Biosignal Interaction and Personhood Technology (BIAPT) Lab, McGill University, Montreal, QC, Canada.
Interpersonal physiological synchrony has been successfully used to characterize social interactions and social processes during a variety of interpersonal interactions. There are a handful of measures of interpersonal physiological synchrony, but those that exist have only been validated on able-bodied adults. Here, we present a novel information-theory based measure of interpersonal physiological synchrony-normalized Symbolic Transfer Entropy (NSTE)-and compare its performance with a popular physiological synchrony measure-physiological concordance and single session index (SSI).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!