Purpose: To assess the comparability of ocular biometry measurements and refractive outcomes between ANTERION and IOL Master 700.
Methods: This comparative prospective study was conducted from December 2020 to February 2021. A total of 225 patients who had cataracts were enrolled for the study and different parameters such as anterior keratometry (Steep K, Flat K) with axis, Sim K, central corneal thickness (CCT), aqueous depth (AQD), lens thickness (LT), and axial length (AL), white-to-white (WTW) were evaluated in IOL master 700 first and then with ANTERION. Finally, 203 patients completed the 6-week follow-up and the postoperative refraction was done at the 6 week. To assess the agreement between the devices, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used. To analyze the agreement for postoperative residual refractive error between the two devices, Kappa statistics were used.
Results: The mean difference for steep K, flat K, and Sim K between ANTERION and IOL Master 700 were - 0.18 +/-0.26 D, -0.13+/- 0.28D, -0.15 +/-0.23, respectively. The CCT, ACD, and LT also showed excellent agreement (ICC > 0.9) but the similarity for the keratometry axis was not up to the mark (ICC = 0.794). For postoperative refractive outcomes, the Kappa value was 0.437, indicating moderate agreement.
Conclusion: : ANTERION showed a good agreement for the majority of parameters with IOLMaster 700 in measuring ocular biometry, except for the keratometry. The accuracy of the intraocular lens power calculations was clinically acceptable with both biometers though the IOL power given by ANTERION remained slightly on the hypermetropic side.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9332939 | PMC |
http://dx.doi.org/10.4103/ijo.IJO_2433_21 | DOI Listing |
J Cataract Refract Surg
January 2025
Yokohama Sky Building Eye Clinic, 2-19-12 Takashima, Yokohama, Nishi-ku, Kanagawa, 220-0011, Japan.
Purpose: To compare the refractive accuracy of the Barrett True axial length (BTAL) formula, newly integrated into ARGOS, with that of the Barrett Universal II (BUII) formula calculated using axial length (AL) from IOL Master 700.
Setting: Private clinics in Kanagawa, Japan.
Design: Retrospective observational study.
J Cataract Refract Surg
January 2025
Kepler University Hospital, University Clinic for Ophthalmology and Optometry.
Purpose: To comprehensively evaluate the tilt of the crystalline lens in phakic eyes and its relationship with optical biometry parameters.
Setting: The retrospective analysis was conducted at the Department for Ophthalmology and Optometry, Kepler University Clinic in Linz, Austria.
Design: Lens tilt assessment was performed through a combined analysis of optical biometry (IOL Master 700) and anterior segment optical coherence tomography (AS-OCT) data to understand its relationship.
J Ophthalmol
January 2025
Pediatric Ophthalmology, Shanxi Aier Eye Hospital, Aier Eye Hospital Group, Changsha, China.
To investigate the correlation between corneal biomechanical characteristics and refractive status in adolescents aged 5-13 years. A cross-sectional study involved 339 children aged 5-13 with a spherical equivalent (SE) range from -6.00 to +2.
View Article and Find Full Text PDFFront Artif Intell
January 2025
School of Medicine, Stanford University, Palo Alto, CA, United States.
Given close relationships between ocular structure and ophthalmic disease, ocular biometry measurements (including axial length, lens thickness, anterior chamber depth, and keratometry values) may be leveraged as features in the prediction of eye diseases. However, ocular biometry measurements are often stored as PDFs rather than as structured data in electronic health records. Thus, time-consuming and laborious manual data entry is required for using biometry data as a disease predictor.
View Article and Find Full Text PDFSci Rep
December 2024
Department of Ophthalmology, Rajavithi Hospital and College of Medicine, Rangsit University, Bangkok, Thailand.
Relative anterior microphthalmos (RAM) is a rare ocular condition characterized by disproportionately small anterior segments but normal axial length (corneal diameter < 11 mm and axial length > 20 mm). This study aimed to determine the prevalence of RAM and its association with glaucoma utilizing IOL Master 700 data (Carl Zeiss Meditec, Jena, Germany). A retrospective analysis was conducted of the biometric parameters of 6,407 eyes, and 115 cases of RAM were identified.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!