Human Exposure to Hand-Arm Vibration from a Surgical Drill During Simulated Spine Surgery.

World Neurosurg

Motion Analysis Laboratory, Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota, USA. Electronic address:

Published: August 2022

Objective: During spine surgery, surgeons are exposed to vibrations from surgical drills. Increased exposure to vibration can result in neurologic, vascular, and musculoskeletal impairments. To reduce these risks, occupational health standards have been implemented to limit exposure levels. The purpose of this study was to quantify human exposure to hand-arm vibration from a surgical drill during a simulation of a common spine procedure.

Methods: Vibration measurements were collected during three 30-second surgical trials on a fresh frozen cadaver torso specimen using a standard surgical drill. The daily vibration exposure A(8) was evaluated on the basis of International Organization for Standardization 5349-1, and the occupational health standards, exposure action value, and exposure limit value were calculated.

Results: On the basis of vibration exposure, surgeons in this study reached their exposure limits in 8 minutes to monitor tool usage and 32 minutes to terminate equipment usage.

Conclusions: The results demonstrate that a common surgical drill transmits hand-arm vibration levels approaching the exposure action value and exposure limit value over short periods of use. Further investigation is needed to determine the total vibration exposure over an entire workday, specifically in surgeons who perform multiple surgeries within a single day.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2022.04.095DOI Listing

Publication Analysis

Top Keywords

surgical drill
16
hand-arm vibration
12
vibration exposure
12
exposure
11
human exposure
8
exposure hand-arm
8
vibration
8
vibration surgical
8
spine surgery
8
occupational health
8

Similar Publications

Smartphone application-based augmented reality for pre-clinical dental implant placement training: a pilot study.

Oral Maxillofac Surg

January 2025

Research Center for Digital Technologies in Dentistry and CAD/CAM, Department of Dentistry, Faculty of Medicine and Dentistry, Danube Private University, Steiner Landstraße 123, Krems an der Donau, 3500, Austria.

Purpose: Precise implant placement is essential for optimal functional and aesthetic outcomes. Digital technologies, such as computer-assisted implant surgery (CAIS), have improved implant outcomes. However, conventional methods such as static and dynamic CAIS (dCAIS) require complex equipment.

View Article and Find Full Text PDF

The endoscopic transsphenoidal approach is a common approach used in skull base neurosurgery to reach the sellar region. One of the intraoperative risks of this approach is intraoperative bleeding out of the carotid artery. Gentle drilling can prevent carotid artery injury.

View Article and Find Full Text PDF

Introduction: Numerous orthopaedic procedures including dynamic hip screw plating and various osteotomies require placement of a reference guide pin or K wire to direct bone cuts or for drilling screw holes. Appropriate positioning of these wires is a critical component of surgery. Irrespective of whether one is a seasoned surgeon or an apprentice, these wires often need repositioning and readjustment.

View Article and Find Full Text PDF

The emergence of augmented reality (AR) in surgical procedures could significantly enhance accuracy and outcomes, particularly in the complex field of orthognathic surgery. This study compares the effectiveness and accuracy of traditional drilling guides with two AR-based navigation techniques: one utilizing ArUco markers and the other employing small-workspace infrared tracking cameras for a drilling task. Additionally, an alternative AR visualization paradigm for surgical navigation is proposed that eliminates the potential inaccuracies of image detection using headset cameras.

View Article and Find Full Text PDF

Effect of Laminectomy Methods on the Surgical Safety of Automatic Laminectomy Robot.

Int J Med Robot

February 2025

Beijing Key Laboratory for Design and Evaluation Technology of Advanced Implantable & Interventional Medical Devices, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China.

Background: The efficacy of laminectomy procedures is contingent on the method of resection. The objective of this study was to investigate the impact of different methods of resection on the surgical safety of automated laminectomy robots, an area that remains uncharted.

Methods: Lamina resection surgeries using both drilling and layer-by-layer methods, are performed on ovine spinal samples.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!