Objectives: The aim of this study was to evaluate outcomes of real-world patients with aortic stenosis (AS) undergoing transcatheter aortic valve replacement (TAVR) with the 34 mm Evolut R (Medtronic, Minneapolis, Minnesota).
Background: Larger aortic annulus has been associated with increased incidence of paravalvular leaks (PVLs) after TAVR. However, little is known, so far, about the performance of the 34 mm Evolut R in this setting.
Methods: From the multicenter, international, retrospective Horizontal Aorta in Transcatheter Self-expanding Valves (HORSE) registry, including patients who underwent TAVR for native severe AS, we selected patients treated with the 34 mm Evolut R evaluating procedural characteristics and VARC-2 defined device success. We also compared 34 mm Evolut R with other Evolut R sizes.
Results: Among the 4434 patients included in the registry, 572 (13%) received the 34 mm Evolut R valve. Mean age was 80.8 ± 6.5 years and the median STS PROM score was 4 [interquartile range 2-6]. Device success was achieved in 87.4% with 7.7% of PVLs; moreover, the rate of permanent pacemaker implantation (PPMI) was 22.4%. Patients who underwent 34 mm Evolut R implantation experienced more in-hospital permanent pacemaker implantation (22.4% vs. 15%; p < 0.001). At multivariate analysis, 34 mm Evolut R did not affect device success (OR: 0.81 [0.60-1.09]; p = 0.151). Device success was consistent with other THVs sizes (87.4% vs. 89.6%; p = 0.157).
Conclusions: THV replacement in patients requiring 34 mm Evolut R has an acceptable performance. Compared to other Medtronic sizes it demonstrated to be comparable in terms of device success, despite an increased rate of pacemaker implantation.