Purpose: Primary or recurrent urethral stenoses are common clinical problems. The aim of this feasibility study was development and application test of a novel radioactive catheter for potential use in low-dose-rate (LDR) brachytherapy to prevent urethral stenosis.

Material And Methods: A beta radiation emitting phosphorous-32 (P)-foil was developed and integrated in indwelling catheter, which is usually inserted after surgical intervention in the urethra. Activity and dose distribution were measured after neutron activation of the phosphorus-31 (P)-foil component with scintillation techniques as well as radiochromic films, and compared to MC simulations. Practicability and radiation safety of this new applicator were tested on male rabbits, which, before catheter application, had undergone a new method of stricture induction by radial laser light irradiation.

Results: Measured dose distributions of catheters with P-foils were found to be in good agreement with MC simulations. Wash-out tests confirmed the radioactive catheter setup to be inside permitted limits of a sealed source. Strictures could be induced by laser light in all animals, surgically treated by slitting and irradiation with a 7-day dose of 15 Gy or 30 Gy, without adverse effects of radiation during a 4-week observation time.

Conclusions: This proof-of-concept study presents a safe and technically simple approach to LDR brachytherapy intending to prevent urethral stenosis. The prescribed irradiation dose can be administered in a reproducible and precise manner using a urethral catheter as a carrier of beta radiating foil. The study duration of 63 days was too short to make statistically significant stricture-related conclusions. This approach may be considered for similar problems in other hollow organs and in cases regarding superficial hyper-proliferation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9044302PMC
http://dx.doi.org/10.5114/jcb.2022.115205DOI Listing

Publication Analysis

Top Keywords

radioactive catheter
8
ldr brachytherapy
8
prevent urethral
8
laser light
8
urethral
5
catheter
5
development pre-clinical
4
pre-clinical test
4
test phosphorous-32
4
phosphorous-32 polyetheretherketone
4

Similar Publications

High dose-rate brachytherapy presents a promising therapeutic avenue for prostate cancer management, involving the temporary implantation of catheters which deliver radioactive sources to the cancerous site. However, as catheters puncture and penetrate the prostate, tissue deformation is evident which may affect the accuracy and efficiency of the treatment. In this work, a data-driven modelling procedure is proposed to simulate brachytherapy while accounting for prostate biomechanics.

View Article and Find Full Text PDF

Background: Comprehensive quality assurance (QA) for a seamless workflow of high-dose-rate brachytherapy, from imaging to planning and irradiation, is uncommon, and QA of the source dwell position is performed in one- or two-dimensions. Gel dosimetry using magnetic resonance imaging (MRI) is effective in verifying the three-dimensional distribution of doses for image-guided brachytherapy (IGBT). However, MRI scanners are not readily accessible, and MRI scanning is time-consuming.

View Article and Find Full Text PDF

Image-guided patient-specific optimization of catheter placement for convection-enhanced nanoparticle delivery in recurrent glioblastoma.

Comput Biol Med

September 2024

Oden Institute for Computational Engineering and Sciences, The University of Texas at Austin, Austin, TX, 78712, USA; Department of Biomedical Engineering, The University of Texas at Austin, Austin, TX, 78712, USA; Department of Diagnostic Medicine, The University of Texas at Austin, Austin, TX, 78712, USA; Department of Oncology, The University of Texas at Austin, Austin, TX, 78712, USA; Livestrong Cancer Institutes, The University of Texas at Austin, Austin, TX, 78712, USA; Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, 77030, USA.

Background: Proper catheter placement for convection-enhanced delivery (CED) is required to maximize tumor coverage and minimize exposure to healthy tissue. We developed an image-based model to patient-specifically optimize the catheter placement for rhenium-186 (Re)-nanoliposomes (RNL) delivery to treat recurrent glioblastoma (rGBM).

Methods: The model consists of the 1) fluid fields generated via catheter infusion, 2) dynamic transport of RNL, and 3) transforming RNL concentration to the SPECT signal.

View Article and Find Full Text PDF

Lung cancer is the leading cause of cancer death worldwide. It may present as airway obstruction in a patient with endobronchial masses. Endobronchial brachytherapy (EBBT) has been shown to provide palliative therapy.

View Article and Find Full Text PDF

Extravasation of the radiopharmaceutical during peptide receptor radionuclide therapy infusion is an unwanted infrequently reported event. We present the case of a 74-year old woman with a neuroendocrine tumor who was referred for peptide receptor radionuclide therapy. During intravenous infusion of 7.

View Article and Find Full Text PDF