Background: Home monitoring devices have been developed to measure adherence to home oxygen therapy. In this study, we evaluated the performance of TeleOx, a commercially available remote monitoring device, in comparison with polysomnography (PSG) in patients with interstitial lung disease (ILD) and chronic obstructive pulmonary disease (COPD) and the factors that affected TeleOx correct use.
Methods: TeleOx was connected on the patient or concentrator side. The oxygen flow rates were set at 1, 3, and 5 L/min. Intraclass correlation coefficient (ICC) (2,1) was used to determine the agreement between respiratory rate measured using TeleOx and that measured using PSG, and the minimum acceptable level of reliability was >0.7.
Results: In total, 22 patients (16 with ILD and 6 with COPD) were assessed. In patients with ILD, the detection rate of patients' respiration assessed using TeleOx did not change according to the device's position. It increased from 53.5% to 79.0% by changing the position from the concentrator to the patient side in patients with COPD. The ICC (2,1) value indicated that TeleOx had acceptable reliability at oxygen flow rates of 1 and 3 L/min regardless of the device's position in patients with ILD (the concentrator side: 0.9 and 0.82, respectively; the patient side: 0.95 and 0.82, respectively), whereas that did only at the oxygen flow rate of 1 L/min and in connecting TeleOx on the patient side in patients with COPD (0.73).
Conclusion: The monitoring performance of TeleOx differed according to its position, oxygen flow rates, and patients' diseases.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9007031 | PMC |
http://dx.doi.org/10.36141/svdld.v39i1.12133 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!