Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year.

Acta Orthop

Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg; Department of Orthopedics, Sahlgrenska University Hospital, Gothenburg, Sweden.

Published: April 2022

AI Article Synopsis

Article Abstract

Background And Purpose: Computed tomography micromotion analysis (CTMA) can be used to determine implant micro-movements using low-dose CT scans. By using CTMA, a non-invasive measurement of joint implant movement is enabled. We evaluated the precision of CTMA in measuring early cup migration. Standard marker-based radiostereometric analysis (RSA) was used as reference. We hypothesised that CTMA can be used as an alternative to RSA in assessing implant micromotions.

Patients And Methods: We included 30 patients undergoing total hip arthroplasty (THA). Acetabular cup migration at 1 year was measured with RSA and CTMA. To determine the precision of both methods, 20 double examinations (postoperatively) with repositioning of the patients were performed. The precision was calculated from zero by assuming that there was no motion of the prosthesis between the 2 examinations.

Results: The precision of RSA ranged from 0.06 to 0.15 mm for translations and 0.21° to 0.63° for rotations. Corresponding values for CTMA were 0.06 to 0.13 mm and 0.23° to 0.35°. A good level of agreement was found between the methods regarding cup migration and rotation at 1 year.

Interpretation: The precision of CTMA in measuring acetabular cup migration and rotation is comparable to marker-based RSA. CTMA could possibly thus be used as an alternative method to detect early implant migration.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9047498PMC
http://dx.doi.org/10.2340/17453674.2022.2528DOI Listing

Publication Analysis

Top Keywords

cup migration
20
acetabular cup
12
micromotion analysis
8
ctma
8
ctma determine
8
precision ctma
8
ctma measuring
8
ctma alternative
8
rsa ctma
8
migration rotation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!