In-Vitro Comparison of Single Limb and Dual Limb Circuit for Aerosol Delivery via Noninvasive Ventilation.

Respir Care

Division of Respiratory Care, Department of Cardiopulmonary Sciences, Rush University, Chicago, Illinois.

Published: July 2022

Background: The effect of single- and dual-limb circuits on aerosol delivery during noninvasive ventilation (NIV) in adult models is unclear.

Methods: A noninvasive ventilator equipped with a single-limb circuit or an ICU ventilator equipped with a dual-limb circuit was connected to a simulated lung. Ventilator parameters were adjusted to maintain a tidal volume at ∼500 mL. Aerosol deposition with different placements of a vibrating mesh nebulizer and humidification conditions were compared. Additional experiments by using a non-vented mask or a vented mask were compared in the single-limb circuit only. Aerosol was collected by a disposable filter placed between the simulated lung and the head model (n = 3), and measured by ultraviolet spectrophotometry (276 nm).

Results: The aerosol deposition varied between 4.12 ± 0.22% and 20.75 ± 0.95%. The greatest aerosol delivery during NIV when using a non-vented mask was found when a vibrating mesh nebulizer was placed between the mask and 15 cm from the exhalation port in the humidified single-limb circuit, and 15 cm from the Y-piece in the inspiratory limb of the humidified dual-limb circuit, and no significant difference of aerosol deposition was found between the two optimal positions (20.03 ± 1.48% vs 18.04 ± 0.93%, respectively; =.042). There was no difference of aerosol delivery in dry versus humidified circuits, except when a vibrating mesh nebulizer was placed at the humidifier inlet in a dual-limb circuit. When using a vented mask, the aerosol deposition was poor (6.56 ± 0.41 ∼ 8.02 ± 0.39%), regardless of vibrating mesh nebulizer positions and humidification types.

Conclusions: During NIV, the aerosol delivery was optimal when a vibrating mesh nebulizer was placed between the non-vented mask and 15 cm from the exhalation port in the single-limb circuit or 15 cm from the Y-piece in the inspiratory limb of the dual-limb circuit; no significant difference was found between the two optimal placements. Humidification had little effect on aerosol delivery. Aerosol delivery was poor in the single-limb circuit with a vented mask.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.09543DOI Listing

Publication Analysis

Top Keywords

aerosol delivery
28
single-limb circuit
20
vibrating mesh
20
mesh nebulizer
20
dual-limb circuit
16
aerosol deposition
16
aerosol
12
non-vented mask
12
vented mask
12
circuit
10

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!