Background: While continuous glucose monitoring (CGM) systems allow precise and real-time blood glucose control, current electrochemicalbased CGM technologies inherently harbor enzyme instability issues. The direct electron transfer (DET) type open circuit potential (OCP) based enzyme sensing principle can minimize the catalytic turnover of the enzyme reaction, thereby providing longer-term operational stability in future CGM glucose sensors.

Method: DET-type OCP based glucose sensors were constructed using gold disk electrodes with glucose dehydrogenase capable of DET which was immobilized using a self-assembled monolayer (SAM). The single enzyme layer prepared on the gold electrode was operated in the presence of glucose, using in vitro buffer solution, continuously for over 3 months with the OCP sensor signal monitored every 10 seconds at 25°C.

Results: The DET-type OCP glucose sensor was continuously operated for more than 3 months without a significant decrease of the sensor signal and sensitivity (slope). These results suggest that the DET-type OCP glucose sensor is far more stable than the sensor constructed based on the amperometric principle. The long-term stability of DET-type OCP glucose sensor is attributed to the enzyme's minimized catalytic reaction during the operation, thereby extending the lifetime of enzyme.

Conclusion: The DET-type OCP glucose sensor can be continuously operated for more than 3 months at 25 °C, in vitro without significant decreases in sensor signal and sensitivity. While the further investigation will be required for in vivo validation, the DET-type OCP glucose sensor is ideal for next generation CGM's, especially in long duration implantable use cases.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9445357PMC
http://dx.doi.org/10.1177/19322968221092449DOI Listing

Publication Analysis

Top Keywords

glucose sensor
24
det-type ocp
24
ocp glucose
20
glucose
12
sensor signal
12
sensor
11
direct electron
8
electron transfer
8
type open
8
open circuit
8

Similar Publications

Glucose Sensor Design Based on Monte Carlo Simulation.

Biosensors (Basel)

January 2025

Faculty of Civil Engineering and Mechanics, Kunming University of Science and Technology, Kunming 650500, China.

Continuous glucose monitoring based on the minimally invasive implantation of glucose sensor is characterized by high accuracy and good stability. At present, glucose concentration monitoring based on fluorescent glucose capsule sensor is a new development trend. In this paper, we design a fluorescent glucose capsule sensor with a design optimization study.

View Article and Find Full Text PDF

Electrochemical Glucose Sensor Based on Dual Redox Mediators.

Biosensors (Basel)

December 2024

Cofoe Medical Technology Co., Ltd., No. 816 Zhenghua Road, Changsha 410021, China.

Electrochemical glucose sensor holds significant promise for the monitoring of blood glucose levels in diabetic patients. In this study, we proposed a novel electrochemical glucose sensor based on 1,10-Phenanthroline-5,6-dione (PD)/Ru(III) as a dual redox mediator. The synergistic effect of PD and Ru(III) was utilized to efficiently facilitate the electron transfer between the enzyme-active center and the electrode.

View Article and Find Full Text PDF

Nutritional intake is closely linked to gonadal development, although the mechanisms by which food intake affects gonadal development are not fully understood. Cholecystokinin (CCK) is a satiety neuropeptide derived from the hypothalamus, and the present study observed that hypothalamic CCK expression is significantly influenced by food intake, which is mediated through blood glucose levels. Interestingly, CCK and its receptors were observed to exhibit a high expression in the hypothalamus-pituitary-gonad (HPG) axis of grass carp (), suggesting that CCK is potentially involved in regulating fish reproduction through the HPG axis.

View Article and Find Full Text PDF

Background: Continuous glucose monitoring (CGM) improves glycemic control and quality of life. Data on glycemic indices and fear of hypoglycemia (FoH) in newly diagnosed T1DM patients are limited.

Aim: To assess the impact of initiating intermittently scanned CGM (isCGM) within 1-6 months of diagnosis on glycemic control and FoH in adults with T1DM.

View Article and Find Full Text PDF

Intrinsic fluorescence hydrogels for ON/OFF screening of antidiabetic drugs: assessing α-glucosidase inhibition by acarbose.

J Mater Chem B

January 2025

Departamento de Química Inorgánica, Orgánica y Bioquímica, Facultad de Ciencias y Tecnologías Químicas, Universidad de Castilla-La Mancha, 13071 Ciudad Real, Spain.

Diabetes remains one of the most prevalent chronic diseases globally, significantly impacting mortality ratetables. The development of effective treatments for controlling glucose level in blood is critical to improve the quality of life of patients with diabetes. In this sense, smart optical sensors using hydrogels, responsive to external stimuli, have emerged as a revolutionary approach to diabetes care.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!