Purpose: Although flash radiation therapy (FLASH-RT) is a promising novel technique that has the potential to achieve a better therapeutic ratio between tumor control and normal tissue complications, the ultrahigh pulsed dose rates (UHPDR) mean that experimental dosimetry is very challenging. There is a need for real-time dosimeters in the development and implementation of FLASH-RT. In this work, we characterize a novel plastic scintillator capable of temporal resolution short enough (2.5 ms) to resolve individual pulses.

Methods: We characterized a novel plastic dosimeter for use in a linac converter to deliver 16-MeV electrons at 100-Gy/s UHPDR average dose rates. The linearity and reproducibility were established by comparing relative measurements with a pinpoint ionization chamber placed at 10-cm water-equivalent depth where the electrometer is not saturated by the high dose per pulse. The accuracy was established by comparing the plastic scintillator dose measurements with EBT-XD Gafchromic radiochromic films, the current reference dosimeter for UHPDR. Finally, the plastic scintillator was compared against EBT-XD films for online dosimetry of two in vitro experiments performed at UHPDR.

Results: Relative ion chamber measurements were linear with plastic scintillator response within ≤1% over 4-20 Gy and pulse frequencies (18-180 Hz). When characterized under reference conditions with NIST-traceability, the plastic scintillator maintained its dose response under UHPDR conditions and agreed with EBT-XD film dose measurements within 4% under reference conditions and 6% for experimental online dosimetry.

Conclusion: The plastic scintillator shows a linear and reproducible response and is able to accurately measure the radiation absorbed dose delivered by 16-MeV electrons at UHPDR. The dose is measured accurately in real time with a greater level of precision than that achieved with a radiochromic film.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15671DOI Listing

Publication Analysis

Top Keywords

plastic scintillator
28
novel plastic
12
dose
9
plastic
8
online dosimetry
8
dose rates
8
16-mev electrons
8
established comparing
8
dose measurements
8
reference conditions
8

Similar Publications

(1) Background: Ultra-high dose rate (UHDR) radiation therapy needs a reliable dosimetry solution and scintillation detectors are promising candidates. In this study, we characterized an inorganic powder-based scintillation detector under a 9 MeV UHDR electron beam. (2) Methods: A mixture of ZnS:Ag powder and optic glue was coupled to an 8 m Eska GH-4001-P polymethyl methacrylate (PMMA) optical fiber.

View Article and Find Full Text PDF

Design optimization of a 1-D array of stemless plastic scintillation detectors.

Med Phys

January 2025

Department of Physics and Atmospheric Science, Dalhousie University, Halifax, Nova Scotia, Canada.

Background: A stemless plastic scintillation detector (SPSD) is composed of an organic plastic scintillator coupled to an organic photodiode. Previous research has shown that SPSDs are ideally suited to challenging dosimetry measurements such as output factors and profiles in small fields. Lacking from the current literature is a systematic effort to optimize the performance of the photodiode component of the detector.

View Article and Find Full Text PDF

Synchrotron microbeam radiotherapy (MRT), which has entered the clinical transfer phase, requires the development of appropriate quality assurance (QA) tools due to very high dose rates and spatial hyperfractionation. A microstrip plastic scintillating detector system with associated modules was proposed in the context of real-time MRT QA. A prototype of such a system with 105 scintillating microstrips was developed and tested under MRT conditions.

View Article and Find Full Text PDF

Solvent-Tuned Plasticity for Various Binder-Free Applications of a New Lead-Free Manganese Halide.

Adv Mater

December 2024

Division of Materials Science and Engineering, Hanyang University, 222 Wangsimni-ro, Seongdong-gu, Seoul, 04763, Republic of Korea.

The development of efficient color conversion layers for μ-LED technology faces significant challenges owing to the limitations of materials that require binders. Binders are typically used to ensure uniform film formation in color-conversion layers, but they often cause optical losses, increase layer thickness, and introduce long-term stability issues. To address the limitations of materials requiring binders, cyclopropyltriphenylphosphonium manganese tetrabromide (CPTPMnBr) is synthesized, a novel lead-free metal halide.

View Article and Find Full Text PDF

Pulse pile-up presents a significant challenge in nuclear radiation measurements, particularly in neutron-gamma pulse shape discrimination, as it causes pulse distortion and diminishes identification accuracy. To address this, we propose an optimized Support Vector Regression (SVR) algorithm for correcting pulse pile-up. Initially, the Dung Beetle Optimizer (DBO) and Whale Optimization Algorithm (WOA) are integrated to refine the correction process, with performance evaluated using charge comparison methods (CCM) for pulse shape discrimination.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?