Purpose: To develop an online graphic processing unit (GPU)-accelerated Monte Carlo-based adaptive radiation therapy (ART) workflow for pencil beam scanning (PBS) proton therapy to address interfraction anatomical changes in patients treated with PBS.
Methods And Materials: A four-step workflow was developed using our in-house developed GPU-accelerated Monte Carlo-based treatment planning system to implement online Monte Carlo-based ART for PBS. The first step conducts diffeomorphic demon-based deformable image registration (DIR) to propagate contours on the initial planning CT (pCT) to the verification CT (vCT) to form a new structure set. The second step performs forward dose calculation of the initial plan on the vCT with the propagated contours after manual approval (possible modifications involved). The third step triggers a reoptimization of the plan depending on whether the verification dose meets the clinical requirements or not. A robust evaluation will be done for both the verification plan in the second step and the reopotimized plan in the third step. The fourth step involves a two-stage (before and after delivery) patient-specific quality assurance (PSQA) of the reoptimized plan. The before-delivery PSQA is to compare the plan dose to the dose calculated using an independent fast open-source Monte Carlo code, MCsquare. The after-delivery PSQA is to compare the plan dose to the dose recalculated using the log file (spot MU, spot position, and spot energy) collected during the delivery. Jaccard index (JI), dice similarity coefficients (DSCs), and Hausdorff distance (HD) were used to assess the quality of the propagated contours in the first step. A commercial plan evaluation software, ClearCheck™, was integrated into the workflow to carry out efficient plan evaluation. 3D Gamma analysis was used during the fourth step to ensure the accuracy of the plan dose from reoptimization. Three patients with three different disease sites were chosen to evaluate the feasibility of the online ART workflow for PBS.
Results: For all three patients, the propagated contours were found to have good volume conformance [JI (lowest-highest: 0.833-0.983) and DSC (0.909-0.992)] but suboptimal boundary coincidence [HD (2.37-20.76 mm)] for organs-at-risk. The verification dose evaluated by ClearCheck™ showed significant degradation of the target coverage due to the interfractional anatomical changes. Reoptimization on the vCT resulted in great improvement of the plan quality to a clinically acceptable level. 3D Gamma analyses of PSQA confirmed the accuracy of the plan dose before delivery (mean Gamma index = 98.74% with a threshold of 2%/2 mm/10%), and after delivery based on the log files (mean Gamma index = 99.05% with a threshold of 2%/2 mm/10%). The average time cost for the complete execution of the workflow was around 858 s, excluding the time for manual intervention.
Conclusion: The proposed online ART workflow for PBS was demonstrated to be efficient and effective by generating a reoptimized plan that significantly improved the plan quality.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.15678 | DOI Listing |
PLoS One
January 2025
Colorectal Cancer Center, Kyungpook National University Chilgok Hospital, School of Medicine, Kyungpook National University, Daegu, Republic of Korea.
This study aimed to identify radiotherapy dosimetric parameters related to local failure (LF)-free survival (LFFS) in patients with lung and liver oligometastases from colorectal cancer treated with stereotactic body radiotherapy (SBRT). We analyzed 75 oligometastatic lesions in 55 patients treated with SBRT between January 2014 and December 2021. There was no constraint or intentional increase in maximum dose.
View Article and Find Full Text PDFCurr Oncol
November 2024
Department of Radiology and Nuclear Medicine, University Medical Center Utrecht, 3584 CX Utrecht, The Netherlands.
In same-day radioembolization, 99mTc-MAA SPECT/CT, 90Y radioembolization, and post-treatment 90Y SPECT/CT procedures are conducted on the same-day, resulting in a dual-isotope environment of 90Y and 99mTc during post-treatment imaging. This study aimed to quantify the impact of 99mTc on 90Y bremsstrahlung-SPECT/CT image quality and to establish an optimised imaging protocol for both clinical practice, and with advanced reconstruction techniques. Utilising a NEMA IQ phantom, contrast recovery coefficients (CRCs) were measured to evaluate the 90Y image quality degradation caused by 99mTc.
View Article and Find Full Text PDFPsychol Methods
December 2024
Brain Research Institute, University of California, Los Angeles.
J Biomech Eng
February 2025
Department of Biomedical Engineering and Science, Florida Institute of Technology, 150 West University Boulevard, Melbourne, FL 32901.
Lipid-rich atheromas are linked to plaque rupture in stented atherosclerotic arteries. While fibrous cap thickness is acknowledged as a critical indicator of vulnerability, it is likely that other morphological features also exert influence. However, detailed quantifications of their contributions and intertwined effects in stenting are lacking.
View Article and Find Full Text PDFMar Pollut Bull
December 2024
Trakya University, İpsala Vocational School, Department of Laboratory Technology, Evrenos Gazi Campus, Edirne, Türkiye.
This study evaluates the nutritional value and health risks of fish from Miliç Wetland, Türkiye, focusing on potentially toxic elements (PTEs) in Esox lucius, Squalius cephalus, and Carassius gibelio. Using ICP-MS, mean PTE concentrations were determined, including Zn (4979 μg/kg), Fe (4241 μg/kg), and As (125 μg/kg). Macro elements like K, P, and Ca were also assessed for nutritional profiling.