Background: The widely automated method using indirect ion specific electrodes (ISE) potentiometry for determination of sodium concentration is prone to interference from lipaemia. Manufacturer-specified lipaemic (L)-index cut offs may underestimate the effects of endogenous lipaemia.
Methods: We assessed the interference on sodium concentration caused by endogenous lipaemia in 32 residual samples (from 13 patients) using indirect ISE (Cobas 8000 modular analyser with c702 module, Roche diagnostics) and direct ISE (GEM 4000 premier, Werfen) potentiometric methods. Regression analysis (linear and non-linear) was used to determine a reliable (L)-index cut off for reporting sodium concentration.
Results: There was a poor correlation observed between triglyceride concentration and (L)-index. There was significant negative interference caused by endogenous lipaemia within analysed samples. Non-linear regression demonstrated a negative interference of approximately 5% at an (L)-index of 250.
Conclusion: At present, the manufacturer advises not to report sodium concentration by indirect ISE on the Cobas 8000 modular analyser if the (L)-index is >2000. However, this has been determined by the addition of exogenous lipids (Intralipid) and it is clear that this is not comparable to endogenous lipaemia. To ensure patient safety, clinical laboratories should consider lowering the cut off for (L)-index that they use for reporting sodium concentration.