Background: Recent research on photon detection has led to the introduction of a silicone photomultiplier (SiPM) that operates at a low voltage and is insensitive to magnetic fields.

Objective: This work aims to model a scintillation camera with a SiPM sensor and to evaluate the camera reconstructed images from gamma ray projection data.

Material And Methods: The type of study in this research is experimental work and analytical. The scintillation camera, modelled from an SiPM sensor array SL4-30035, coupled with a scintillation material Caesium Iodide doped with Thallium (CsI(Tl)), is used in the experimental part. The performance of the camera was evaluated from reconstructed images by a back-projection technique of a radioactive source Caesium-137 (Cs-137).

Results: The experiments conducted with a 1 µCi Cs-137 radioactive source have revealed that the bias voltage ( ) of the SiPM needs to be set to 27.8 V at an operating temperature between 43 °C to 44 °C. The radioactive source has to be placed within a 1 cm distance from the sensor to obtain the optimum projection data. Finally, the back-projection technique for image reconstruction with linear interpolation pre-processing has revealed that the Ram-Lak filter produces a better image contrast ratio compared to others.

Conclusion: This research has successfully modelled a scintillation camera with SiPM that was able to reconstruct images with an 86.4% contrast ratio from gamma ray projection data.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8995755PMC
http://dx.doi.org/10.31661/jbpe.v0i0.2009-1183DOI Listing

Publication Analysis

Top Keywords

scintillation camera
16
radioactive source
12
image reconstruction
8
silicone photomultiplier
8
camera sipm
8
sipm sensor
8
reconstructed images
8
gamma ray
8
ray projection
8
back-projection technique
8

Similar Publications

First Retrospective QA of FAST-01 Clinical Treatment Fields Using High-Speed Quantitative Scintillation Imaging.

Int J Radiat Oncol Biol Phys

December 2024

Cancer and Blood Diseases Institute, Cincinnati Children's Hospital, Cincinnati, Ohio; Department of Radiation Oncology, College of Medicine, University of Cincinnati, Cincinnati, Ohio.

Purpose: To retrospectively validate the dose and dose rates delivered in XXX clinical trial fields via sub-millimeter spatial and <0.25 ms temporal resolution scintillation imaging.

Methods: An ultra-fast intensified CMOS camera (4.

View Article and Find Full Text PDF

Novel photopeak-independent correction method for internal activity calculation of 99mTc: a simulation study.

Radiat Prot Dosimetry

December 2024

Facultad de Medicina, Universidad Autónoma del Estado de México, Paseo Tollocan s/n esquina Jesús Carranza, Col. Moderna de la Cruz, Toluca 50180, Estado de México, Mexico.

This paper presents a new method for correcting the contribution of scattered radiation to the measurement of 99mTc internal activity in nuclear medicine patients using gamma cameras. So, this study aims to derive scattering correction factors by Monte Carlo simulation for anterior and posterior count rates (${I}_{\mathrm{A}}$ and ${I}_{\mathrm{P}}$) in the conjugate view method, enabling more precise estimation of activity A(t) compared to traditional trapezoidal and triangular approximations. The new approach eliminates the need to use photopeak for determining the fraction of scattered photons.

View Article and Find Full Text PDF

Recent Progress in Myocardial Perfusion Imaging Techniques.

Nucl Med Mol Imaging

December 2024

Department of Nuclear Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Article Synopsis
  • Nuclear cardiology has advanced rapidly, particularly in myocardial perfusion imaging (MPI) since the 1970s, which is vital for assessing and managing coronary artery disease.
  • Recent innovations include the development of gamma cameras with cadmium zinc telluride (CZT) systems and enhancements in software, as well as ongoing studies into new radiopharmaceuticals for positron emission tomography (PET)-MPI.
  • Despite these advancements, challenges like high equipment and examination costs, along with limited accessibility, pose significant barriers to the practical use of these technologies in nuclear medicine.
View Article and Find Full Text PDF

Background: Pre-clinical studies demonstrate that delivering a high dose at a high dose rate result in less toxicity while maintaining tumor control, known as the FLASH effect. In proton therapy, clinical trials have started using 250 MeV transmission beams and more trials are foreseen. A novel aspect of FLASH treatments, compared to conventional radiotherapy, is the importance of dose rate next to dose and geometry.

View Article and Find Full Text PDF

Underwater sensors and autonomous underwater vehicles (AUVs) are widely adopted in oceanic research activities. As the number of underwater sensors and AUVs is growing quickly, the bandwidth requirements are increasing accordingly. In this work, we put forward and demonstrate a large field-of-view (FOV) water-to-air unmanned aerial vehicle (UAV) based optical camera communication (OCC) system with gated recurrent unit neural network (GRU-NN) for the first time to the best of our knowledge.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!