Objective: A heightened and organized understanding of sacral anatomy could potentially lead to a more effective and safe method of dorsal root ganglion stimulation (DRG-S) lead placement. The aim of this technical note is to describe a standardized access method for S1 DRG-S lead placement.
Design: Technical note.
Methods: The described approach utilizes alignment of the lumbosacral prominence and is measurement-based, allowing for standardized sacral access, even when visualization is suboptimal. The medial-to-lateral needle trajectory is designed to limit interaction with the sensitive neural structures and allows for a more parallel orientation of the lead to the DRG and nerve root.
Conclusions: The described technique potentially improves the safety of S1 DRG-S lead placement. The parallel lead orientation to the DRG may also increase efficacy while lowering energy requirements.