The ProtekDuo as double lumen return cannula in V-VP ECMO configuration: A first-in-man method description.

Ann Card Anaesth

Nazih Zuhdi Transplant Institute, Advanced Cardiac and Critical Care, 24/7 Shock Service, Intergris Baptist Medical Center, Oklahoma City, OK, USA.

Published: April 2022

We present a case of acute respiratory distress syndrome (ARDS) secondary to COVID-19 who required venovenous extracorporeal membrane oxygenation (V-V ECMO). Initially, a right ventricular assist device (RVAD), the ProtekDuo with an oxygenator, was placed in an outside heart center and the patient was transferred to us for ECMO management. Due to severe hypoxia, the configuration was later modified, and a 25 Fr femoral drainage cannula was inserted for venous drainage only. The arterial return tubing was spliced and using a Y-connector, arterialized blood was returned through both limbs of the ProtekDuo resulting in a significantly increased oxygenation and flow.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9244280PMC
http://dx.doi.org/10.4103/aca.aca_49_21DOI Listing

Publication Analysis

Top Keywords

protekduo double
4
double lumen
4
lumen return
4
return cannula
4
cannula v-vp
4
v-vp ecmo
4
ecmo configuration
4
configuration first-in-man
4
first-in-man method
4
method description
4

Similar Publications

Test Bench for Right Ventricular Failure Reversibility: The Hybrid BiVAD Concept.

J Clin Med

December 2023

Cardiac Surgery Unit, Department of Cardiac, Thoracic, Vascular Sciences and Public Health, University of Padua, 35121 Padua, Italy.

Background: When heart transplantation and myocardial recovery are unlikely, patients presenting with biventricular cardiogenic shock initially treated with extracorporeal membrane oxygenation (ECMO) may benefit from a mechanical support upgrade. In this scenario, a micro-invasive approach is proposed: the combination of the double-lumen ProtekDuo cannula (Livanova, London, UK) and the Impella 5.5 (Abiomed, Danvers, MA) trans-aortic pump that translates into a hybrid BiVAD.

View Article and Find Full Text PDF

This article describes a case study where a patient with ProtekDuo + oxygenator and Impella CP (PROpella) in biventricular failure and cardiogenic shock was reconfigured to venopulmonary-arterial (VP-A) extracorporeal membrane oxygenation (ECMO), secondary to leg ischemia. The ProtekDuo was subsequently used as double-lumen drainage cannula, returning blood into a newly percutaneously placed femoral arterial cannula, mimicking venoarterial (V-A) ECMO in VP-A configuration. The following day, an Impella 5.

View Article and Find Full Text PDF

Objective: Assessment of the results of the ProtekDuo cannula applied for dedicated right ventricular support with oxygenator in ARDS secondary to COVID-19.

Methods: Systematic literature search in NHS library, Medline (Pubmed) and EMBASE using appropriate keywords as well as PICOS and PRISMA approach.

Results: Out of 285 publications found, 5 publications met the search criteria and were included in this review.

View Article and Find Full Text PDF

The ProtekDuo for percutaneous V-P and V-VP ECMO in patients with COVID-19 ARDS.

Int J Artif Organs

December 2022

Nazih Zuhdi Transplant Institute, Advanced Cardiac and Critical Care, 24/7 Shock Service, Intergris Baptist Medical Center, Oklahoma City, OK, USA.

Objective: The ProtekDuo with oxygenator mimics veno-venous (V-V) extracorporeal membrane oxygenation (ECMO) in veno-pulmonary (V-P) configuration. We have recently developed a new configuration by utilizing a 25 Fr multistage femoral venous drainage cannula and by returning oxygenated blood through both lumina of the double lumen ProtekDuo cannula (V-VP configuration), thereby creating partial right ventricular bypass and oxygenated blood flow of up to seven LPM. We investigated our experience with V-P and V-VP ECMO in patients suffering from COVID-19 acute respiratory distress syndrome (ARDS).

View Article and Find Full Text PDF

The ProtekDuo, single site dual lumen cannula can be used in a multitude of configurations. In the medical literature, the ProtekDuo is described as right ventricular assist device (RVAD), as left ventricular assist device (LVAD), as biventricular assist device (BiVAD) when combined with an Impella device or other LVAD's, as double lumen drainage cannula on cardiopulmonary bypass (CPB), as RVAD with oxygenator, in veno-pulmonary (V-P) ECMO configuration, venovenous-pulmonary (VV-P) ECMO configuration with additional drainage cannula, and in veno-venopulmonary (V-VP) ECMO configuration as double lumen return cannula. Improvements in flow and oxygenation have been recognized in various settings and need further scientific evaluation.

View Article and Find Full Text PDF