Purpose: Cryoballoon (CB) ablation and radiofrequency (RF) ablation are the most common techniques for pulmonary vein isolation (PVI) in patients with symptomatic atrial fibrillation (AF). An increasing number of patients undergoing PVI are obese. To address the paucity of data on outcomes of CB- vs. RF-based PVI in relation to body mass index (BMI) of AF patients.

Methods: All patients undergoing de novo PVI between 01/2018 and 08/2019 at University Hospital Cologne were included in this retrospective analysis. Patients of each group (CB-PVI vs. RF-PVI) were analyzed based on their BMI. Hereafter, procedural characteristics and AF recurrence rate were compared regarding different BMI groups.

Results: A total of 526 patients (62% male, 65±11 years) underwent successful de novo PVI (320 CB and 206 RF). In obese patients, two differences in procedural characteristics were noted: A significantly increased contrast medium volume in CB group and a lower fluoroscopy dose in RF group: contrast medium: CB 50 [40-80] vs. RF 20 [20-30], p<0.001; fluoroscopy dose: CB 392.4 [197.9-995.9] vs. RF 282.5 [139.8-507.2], p<0.001. The complication rate was equal throughout all BMI groups, regardless of CB or RF usage. For obese patients, a trend toward a higher AF recurrence rate was revealed after RF-PVI as compared to CB-PVI. In line with previous studies, the overall procedure time was significantly shorter with CB-PVI regardless of BMI.

Conclusion: For obese patients, CB-PVI is similarly safe and effective as RF-PVI. The significantly shorter procedure time for CB-PVI may minimize potential obesity-related complications. However, the lower contrast medium quantity and fluoroscopy dose in RF-PVI must be considered. AF recurrence rates were comparable between CB-PVI and RF-PVI.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9640436PMC
http://dx.doi.org/10.1007/s10840-022-01210-3DOI Listing

Publication Analysis

Top Keywords

pulmonary vein
8
vein isolation
8
atrial fibrillation
8
patients undergoing
8
novo pvi
8
procedural characteristics
8
contrast medium
8
patients
7
pvi
5
procedural outcome
4

Similar Publications

Background: Pulsed field ablation (PFA) is gaining recognition as a nonthermal, tissue-specific technique for the treatment of atrial fibrillation (AF). The preclinical evaluation of the investigated novel PFA system from Insight Medtech Co. Ltd has demonstrated feasibility, safety, and 30-day efficacy for pulmonary vein isolation (PVI) in the swine model.

View Article and Find Full Text PDF

Characteristics of In Vivo Lesion Formation With a Temperature-Controlled Diamond-Tip Radiofrequency Ablation Catheter in the Ventricle: A Preclinical Model.

Circ Arrhythm Electrophysiol

January 2025

Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN (T.H., M.E.R., O.Y., G.N.K., N.O., T.K., L.N., D.L.P., K.C.S.).

Background: Power-controlled radiofrequency ablation with irrigated-tip catheters has been the norm for ventricular ablation for almost 2 decades. New catheter technology has recently integrated more accurate tissue temperature sensing enabling temperature-controlled irrigated ablation. We aimed to investigate the in vivo ablation parameters and lesion formation characteristics in ventricular myocardium using a novel temperature-controlled radiofrequency catheter.

View Article and Find Full Text PDF

Background: Areas of conduction disorders play an important role in both initiation and perpetuation of AF and can be recognized by specific changes in unipolar potential morphology. For example, EGM fractionation may be caused by asynchronous activation of adjacent cardiomyocytes because of structural barriers such as fibrotic strands. However, it is unknown whether there are sex differences in unipolar potential morphology.

View Article and Find Full Text PDF

Remaining of contrast dye ice cap during PVI by cryoballoon ablation; a case report.

J Cardiothorac Surg

January 2025

Department of Cardiology, School of Medicine, Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran.

Pulmonary vein isolation (PVI) with cryoballoon (CB) ablation technology is widely used to treat drug-resistant atrial fibrillation (AF). During CB ablation, there is a possibility of forming an ice cap of contrast-color on top of the balloon. If automatic balloon deflate occurs before the ice cap dissolves, embolization to the systemic circulation is possible.

View Article and Find Full Text PDF

Results of ICE-Guided Isolation of the Superior Vena Cava With Pulsed Field Ablation.

JACC Clin Electrophysiol

January 2025

Texas Cardiac Arrhythmia Institute, St. David's Medical Center, Austin, Texas, USA; Interventional Electrophysiology, Scripps Clinic, San Diego, California, USA; Metro Health Medical Center, Case Western Reserve University School of Medicine, Cleveland, Ohio, USA; Department of Biomedicine and Prevention, Division of Cardiology, University of Tor Vergata, Rome, Italy. Electronic address:

Background: Earlier studies have documented the risk for sinoatrial node injury and phrenic nerve paralysis as complications following radiofrequency catheter ablation for electrical isolation of the superior vena cava (SVCI).

Objectives: The aim of this study was to assess the safety and feasibility of SVCI in patients with atrial fibrillation undergoing pulsed field ablation (PFA) METHODS: Six hundred sixteen consecutive patients undergoing PFA for pulmonary vein isolation plus SVCI were included in this multicenter analysis. Superior vena cava (SVC) ablation was performed under the continuous guidance of intracardiac echocardiography.

View Article and Find Full Text PDF