Purpose: Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities.

Methods: Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples.

Results: We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types ( = .323).

Conclusions: This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs.: Therapeutic III.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991424PMC
http://dx.doi.org/10.1016/j.jhsg.2021.03.004DOI Listing

Publication Analysis

Top Keywords

nitinol staples
20
olecranon osteotomies
12
average displacement
12
displacement inset
12
olecranon osteotomy
8
fixation
8
osteotomy fixation
8
inset
8
staples
8
staples inset
8

Similar Publications

Many fixation options exist for correction of a hallux valgus (HV) deformity with a first tarsometatarsal (TMT) joint arthrodesis with pros and cons to each. Few have looked at utilizing a SMA (shape memory alloy) staple made from nitinol for fixation. This study evaluates the effectiveness of a modified Lapidus technique using a nitinol (shape memory alloy) staple combined with a two-screw construct for first tarsometatarsal (TMT) joint arthrodesis in correcting hallux valgus (HV) deformity.

View Article and Find Full Text PDF

Nitinol Compression Staples in Foot Orthopaedic Surgery: A Systematic Review.

Foot Ankle Orthop

October 2024

Department of Orthopaedic Surgery and Sports Medicine, University of Florida, Gainesville, FL, USA.

Background: Nitinol compression staples have been increasingly used in foot and ankle orthopaedic surgery because of simple implantation, reproducibility, and favorable biomechanical features. Compared with traditional implants (i.e.

View Article and Find Full Text PDF

Background: There are a variety of ways to surgically manage patients with scaphoid waist fractures. The purpose of this study is to compare the rate of union achieved with a nitinol compression staple versus that of a headless compression screw in the treatment of scaphoid waist fractures.

Methods: We performed a retrospective review of patients with middle-third scaphoid fractures treated surgically.

View Article and Find Full Text PDF
Article Synopsis
  • The study offers a background on the subject being investigated.
  • It outlines specific methods used to gather and analyze data.
  • Key results include findings represented by ×, ×, ×, leading to important conclusions about the topic.
View Article and Find Full Text PDF

Forearm diaphyseal fractures are common orthopedic injuries that typically require surgical intervention using various implants and approaches. Maintaining reduction while simultaneously achieving compression in radial and/or ulnar shaft fractures during compression plate application can be challenging, particularly with unstable segmental and/or transverse fracture patterns. Nitinol compression staples have become increasingly used as a reduction aid because of their ability to provide continuous compression between the staple legs at the fracture site, low profile, and ease of application.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!