Introduction: Intramedullary (IM) nailing is a well-established method of treating closed tibia shaft fracture. We hereby describe an unusual incidence of the incarceration of the flexible IM reamer in the distal fracture fragment with breakage of the coiled shaft of the reamer in the proximal third of tibia during closed IM nailing. This case report highlights the usefulness of making a small window in the tibia to aid retrograde removal of incarcerated reamer.

Case Presentation: A 26-year-old male patient had road traffic accident and sustained a close fracture of the right distal 3rd tibia along with segmental fracture of the adjacent fibula. The tibia nailing was done under combined spinal and epidural anesthesia. The proximal tibia was approached by a midline incision with splitting of the patellar tendon. The reduction of the fracture fragments was done and the guide wire was inserted. The medullary canal was narrow and hence the initial reaming was started with the smallest available, that is, 8 mm reamer. After crossing the fracture site, we noticed that reamer coiled shaft got broken into multiple pieces in the medullary cavity at proximal 3rd tibia. The fracture site was exposed and a linear osteotomy was done in distal tibial fragment. Through this window, the broken reamer was pushed in the retrograde manner and was delivered out. Most of the broken metal pieces of the reamer shaft were removed with pituitary rongeur. An 8 mm solid IM nail was passed in antegrade manner across the fracture site and was locked distally and proximally. The operative wounds were irrigated and closure was done in layers. The patient was allowed partial weight bearing at 3 weeks post-surgery with gradually progressing to full weight bearing at 10 weeks after confirming clinical and radiological union.

Conclusion: To handle this unusual occurrence one needs to stay calm, make a small window to aid retrograde removal of reamer and remove the fragmented pieces of the coiled shaft of the reamer. In tight medullary canal, it is desirable to have smaller diameter reamers, that is, 6 mm during the surgery.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8930373PMC
http://dx.doi.org/10.13107/jocr.2021.v11.i12.2582DOI Listing

Publication Analysis

Top Keywords

coiled shaft
16
fracture site
12
reamer
9
unusual occurrence
8
breakage coiled
8
intramedullary nailing
8
tibia
8
fracture
8
shaft reamer
8
small window
8

Similar Publications

Torque measurement is a key task in several mechanical and structural engineering applications. Most commercial torquemeters require the shaft to be interrupted to place the sensors between the two portions of the shaft where a torque has to be measured. Contactless torquemeters based on the inverse magnetostrictive effect represent an effective alternative to conventional ones.

View Article and Find Full Text PDF

Electromagnetic tracking (EMT) can benefit image-guided interventions in cases where line of sight is unavailable. However, EMT can suffer from electromagnetic distortion in the presence of metal instruments. Metal instruments are widely used in laparoscopic surgery, ENT surgery, arthroscopy and many other clinical applications.

View Article and Find Full Text PDF
Article Synopsis
  • The study compared renovated double J stents (DJS) with conventional stents using ureter models to assess their effectiveness for different ureteral diseases.
  • Two ureter models were tested: one with stenosis and one without, using various sizes and designs of DJS to evaluate urine flow.
  • Results showed that renovated stents had better flow rates in models without stenosis but did not perform significantly better than conventional stents in models with stenosis.
View Article and Find Full Text PDF

Background: Endovascular treatment devices require compatible guide catheters to navigate complex vessels and anatomy. The Fubuki XF Long Sheath guide catheter (Fubuki XF) was developed with a 0.090-inch internal diameter with hydrophilic coating, an atraumatic rounded tip, and enhanced trackability and support with gradual shaft transition zones.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!