The novel Eversense XL continuous glucose monitoring system (Senseonics, Inc., Germantown, Maryland) has recently been developed for monitoring diabetes in humans. The sensor is fully implanted and has a functional life of up to 180 days. The present study describes the use of Eversense XL in three diabetic dogs (DD) with good glycemic control managed by motivated owners. The insertion and use of the device were straightforward and well tolerated by the dogs. During the wearing period, some device-related drawbacks, such as sensor dislocation and daily calibrations, were reported. A good correlation between the glucose values measured by the Eversense XL and those obtained with two commercially available devices, previously validated for use in DD, was found (r = 0.85 and r = 0.81, respectively). The life of the sensor was 180 days in two of the DD and provided high satisfaction. This innovative device might be considered a future alternative for home glucose monitoring in DD.