True-positive reflex threshold value for HCV antibody screening test.

J Infect Dev Ctries

Department of Biological Sciences and Biotechnology, Faculty of Science and Technology, The National University of Malaysia, Bangi, Selangor, Malaysia.

Published: March 2022

Introduction: The use of a signal-to-cut-off ratio has been recommended by the Centre for Disease Control and Prevention to determine the need for further validation using a supplemental test. In this study, we aimed to determine the optimal true-positive signal-to-cut-off ratio for the ABBOTT ARCHITECT i2000SR immunoassay (Abbott Laboratories, Illinois, USA), using the Serodia® HCV particle agglutination (HCV-PA) assay (Fujirebio Inc, Tokyo, Japan) as the reference test for anti-HCV screening.

Methodology: We analysed a total of 13,240 specimens using the ARCHITECT i2000SR immunoassay and subsequently subjected all the reactive specimens with a signal-to-cut-off ratio ≥ 1.00 (n = 267) to the Serodia® HCV-PA reference assay. Receiver operating characteristic (ROC) curve analysis was carried out and performance characteristics for each signal-to-cut-off ratio were determined. The selected signal-to-cut-off ratio value was then assessed using a line immunoassay (LIA) test.

Results: ROC curve analysis determined that the optimal signal-to-cut-off ratio was 5.05, which gave the highest Youden's Index (J) value of 0.89, with a sensitivity of 93.1% (88.9-97.2), a specificity of 96.0% (92.4-99.4), a positive predictive value of 96.4% (93.3-99.5), and a negative predictive value of 92.2% (87.5-96.8). Validation of the optimal S/Co value using the LIA test yielded an accuracy of 91.8%, with sensitivity and specificity values of 92.0% and 91.7%, respectively.

Conclusions: The optimal signal-to-cut-off ratio value for the ARCHITECT i2000SR immunoassay, which was determined using HCV-PA assay as the reference test and validated using a HCV-LIA assay, showed high sensitivity and specificity, and may be used in routine anti-HCV screening.

Download full-text PDF

Source
http://dx.doi.org/10.3855/jidc.13750DOI Listing

Publication Analysis

Top Keywords

signal-to-cut-off ratio
28
architect i2000sr
12
i2000sr immunoassay
12
hcv-pa assay
8
reference test
8
roc curve
8
curve analysis
8
optimal signal-to-cut-off
8
sensitivity specificity
8
signal-to-cut-off
7