Objectives: Image quality (IQ) of diffusion-weighted imaging (DWI) with single-shot echo-planar imaging (ssEPI) suffers from low signal-to-noise ratio (SNR) in high b-value acquisitions. Compressed SENSE (C-SENSE), which combines SENSE with compressed sensing, enables SNR to be improved by reducing noise. The aim of this study was to compare IQ and prostate cancer (PC) detectability between DWI with ssEPI using SENSE (EPIS) and using C-SENSE (EPICS).

Materials And Methods: Twenty-five patients with pathologically proven PC underwent multi-parametric magnetic resonance imaging at 3T. DW images acquired with EPIS and EPICS were assessed for the following: lesion conspicuity (LC), SNR, contrast-to-noise ratio (CNR), mean and standard deviation (SD) of apparent diffusion coefficient (ADC) of lesion (lADCm and lADCsd), coefficient of variation of lesion ADC (lADCcv), and mean ADC of benign prostate (bADCm).

Results: LC were comparable between EPIS and EPICS (p > 0.050), and SNR and CNR were significantly higher in EPICS than EPIS (p = 0.001 and p < 0.001). In both EPIS and EPICS, lADCm was significantly lower than bADCm (p < 0.001). In addition, lADCcv was significantly lower in EPICS than in EPIS (p < 0.001).

Conclusion: Compared with EPIS, EPICS has improved IQ and comparable diagnostic performance in PC.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10334-022-01010-wDOI Listing

Publication Analysis

Top Keywords

epis epics
16
single-shot echo-planar
8
diffusion-weighted imaging
8
compressed sense
8
epics epis
8
epis
7
epics
6
clinical application
4
application single-shot
4
echo-planar diffusion-weighted
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!