Prepolarized MRI (PMRI) is a long-established technique conceived to counteract the loss in signal-to-noise ratio (SNR) inherent to low-field MRI systems. When it comes to hard biological tissues and solid-state matter, PMRI is severely restricted by their ultra-short characteristic relaxation times. Here we demonstrate that efficient hard-tissue prepolarization is within reach with a special-purpose 0.26 T scanner designed for ex vivo dental MRI and equipped with suitable high-power electronics. We have characterized the performance of a 0.5 T prepolarizer module, which can be switched on and off in 200 μs. To this end, we have used resin, dental and bone samples, all with times of the order of 20 ms at our field strength. The measured SNR enhancement is in good agreement with a simple theoretical model, and deviations in extreme regimes can be attributed to mechanical vibrations due to the magnetic interaction between the prepolarization and main magnets.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9540585 | PMC |
http://dx.doi.org/10.1002/nbm.4737 | DOI Listing |
Chemphyschem
October 2023
University of North Carolina at Chapel Hill, Chapel Hill, NC, U.S.
Ultra-low field nuclear magnetic resonance spectroscopy (NMR) and imaging (MRI) inherently suffer from a low signal-to-noise ratio due to the small thermal polarization of nuclear spins. Transfer of polarization from a pre-polarized spin system to a thermally polarized spin system via the Spin Polarization Induced Nuclear Overhauser Effect (SPINOE) could potentially be used to overcome this limitation. SPINOE is particularly advantageous at ultra-low magnetic field, where the transferred polarization can be several orders of magnitude higher than thermal polarization.
View Article and Find Full Text PDFJ Magn Reson
March 2023
Univ. Bordeaux, CNRS, CRMSB, UMR 5536, F-33000 Bordeaux, France. Electronic address:
Development of very-low field MRI is an active area of research. It aims at reducing operating costs and improve portability. However, the signal-to-noise issue becomes prominent at ultra-low field (<1 mT), especially for molecular imaging purposes that addresses specific biochemical events.
View Article and Find Full Text PDFAnal Chem
January 2023
Helmholtz-Institut Mainz, GSI Helmholtzzentrum für Schwerionenforschung, Mainz55128, Germany.
Zero- to ultralow-field nuclear magnetic resonance is a modality of magnetic resonance experiment which does not require strong superconducting magnets. Contrary to conventional high-field nuclear magnetic resonance, it has the advantage of allowing high-resolution detection of nuclear magnetism through metal as well as within heterogeneous media. To achieve high sensitivity, it is common to couple zero-field nuclear magnetic resonance with hyperpolarization techniques.
View Article and Find Full Text PDFJ Appl Phys
November 2022
MPA-Quantum, Los Alamos National Laboratory, P.O. Box 1663, MS-D454, Los Alamos, New Mexico 87545, USA.
We present high-resolution magnetic resonance imaging (MRI) at ultra-low field (ULF) with a proton Larmor frequency of around 120 kHz. The key element is a specially designed high-sensitivity sensing coil in the shape of a solenoid with a few millimeter gap between windings to decrease the proximity effect and, hence, increase the coil's quality ( ) factor and sensitivity. External noise is strongly suppressed by enclosing the sensing coil in a copper cylindrical shield, large enough not to negatively affect the coil's factor and sensitivity, measured to be 217 and 0.
View Article and Find Full Text PDFPLoS One
September 2022
Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, Connecticut, United States of America.
Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!