Accuracy of intravitreal injection volume for aflibercept pre-filled syringe and BD Luer-Lok one-milliliter syringe.

Int J Retina Vitreous

Kresge Eye Institute, Department of Ophthalmology, Visual and Anatomical Sciences, Wayne State University School of Medicine, 4717 St. Antoine Street, Detroit, MI, 48201, USA.

Published: April 2022

Background: To evaluate the accuracy of intravitreal injection volume of the pre-filled syringe (PFS) in which aflibercept is packaged compared to the BD Luer-Lok 1-mL syringe.

Methods: Ophthalmologists injected their typical intravitreal volume for aflibercept using either the PFS or BD Luer-Lok 1-mL syringe for 5 times each. The injected fluid was weighed using a micro-scale and converted to volume. The volume of fluid injected was also evaluated when the 0.05 mL line on the PFS was lined up to the tip or base of the dome-shaped plunger.

Results: Injection volume was measured for 12 physicians. The average injected fluid volume was 74.22 ± 15.87 µL for PFS and 53.42 ± 4.61 µL for the BD Luer-Lok 1-mL syringe (p < 0.0001). The average deviation in volume injected for the PFS was higher compared to the BD Luer-Lok 1-mL syringe (11.36 µL vs. 3.35 µL, p < 0.0001). When the PFS was lined up with the tip of the dome-shaped plunger at the 0.05-mL line, the average injected volume was 71.03% higher.

Conclusions: The intravitreal injection volume and variability using the new PFS were significantly higher than the volume injected using the BD Luer-Lok 1-mL syringe previously used, potentially leading to higher rates of visually significant elevation of intraocular pressures.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8981608PMC
http://dx.doi.org/10.1186/s40942-022-00375-3DOI Listing

Publication Analysis

Top Keywords

injection volume
12
luer-lok 1-ml
12
accuracy intravitreal
8
intravitreal injection
8
volume aflibercept
8
pre-filled syringe
8
1-ml syringe
8
injected fluid
8
volume
7
syringe
5

Similar Publications

Purpose: The common approach for organ segmentation in hybrid imaging relies on coregistered CT (CTAC) images. This method, however, presents several limitations in real clinical workflows where mismatch between PET and CT images are very common. Moreover, low-dose CTAC images have poor quality, thus challenging the segmentation task.

View Article and Find Full Text PDF

Introduction: The strategy of adipose component transplantation has been proposed and widely used in both reconstructive and aesthetic surgery. However, there is no uniform standard for the preparation of component fat, and the volume calculation of liposuction and injection in clinical applications is mostly based on experience. This study aims to analyze the volume of component fat obtained during clinical series.

View Article and Find Full Text PDF

Background: Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ).

View Article and Find Full Text PDF

Few studies today address trans-operative analgesia provided by tramadol without local anesthetics for intra-abdominal procedures. The objective of this study was to assess the efficacy of trans-operative analgesia provided by epidurally administered tramadol in cats undergoing elective ovariohysterectomy. For this purpose, 16 healthy queens were randomly assigned to participate in one of two groups: GC, control group, 0.

View Article and Find Full Text PDF

This study investigates the use of nano-sized oxygen bubbles (NOBs) to enhance BTEX (benzene, toluene, ethylbenzene, xylene) biodegradation in groundwater. Optimized NOBs, averaging 155 nm and at a concentration of 6.59 × 10⁸ bubbles/mL, were found to provide sustained oxygen release with a half-life of approximately 50 days.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!