Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 143
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3098
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Severity: Warning
Message: Attempt to read property "Count" on bool
Filename: helpers/my_audit_helper.php
Line Number: 3100
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3100
Function: _error_handler
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The COVID-19 pandemic revealed flaws in the stockpiling and distribution of ventilators. In this study, we assessed the durability, sterilizability, and performance of a 3D-printed ventilator.
Methods: SLS-printed devices were dropped from 1.83 m and autoclaved before evaluation on a COVID-19 simulated patient. The respiratory performance of an extrusion-printed device was studied using a variable compliance model. Ranges of sustainable respiratory rates were evaluated as a function of tidal volume.
Results: Autoclaving and dropping the device did not negatively impact minute ventilation or PIP for sustained ventilation. Equivalence was significant across all measures except for comparing the autoclaved and dropped with p = 0.06. Extrusion produced ventilators achieved minute ventilation ranging from 4.1 to 12.2 L/min for all simulated compliances; there was an inverse correlation between tidal volume and respiratory rate.
Conclusion: The CRISIS ventilator is a durable, sterilizable, and reusable 3D-printed ventilator using off-the-shelf materials which could be employed variety of adult lung diseases. Further in-vivo testing is needed.