Background: Radiation exposure and imaging quality are among the main concerns in endovascular procedures. The Clear VD11 PURE platform technology system (Siemens Healthineers, Erlangen, Germany) has been reported to lower the radiation dose and improve image quality. In the present study, we evaluated whether the radiation dose during peripheral arterial endovascular procedures had decreased after implementation of this new imaging system.

Methods: The patient characteristics (age, gender, body mass index [BMI]), procedure type (diagnostic, balloon angioplasty, atherectomy, stenting), body location (aortoiliac, superficial femoral artery, tibial artery), reference air kerma (RAK), kerma area product (KAP), and fluoroscopy time (FT) were recorded during peripheral artery interventions performed 1 year before (group A) and 1 year after (group B) the CLEAR system upgrade. The procedures were performed in an Artis zeego hybrid room (Siemens Healthineers) with the same providers. A general linear model was used to estimate the average difference between groups adjusted by procedure type and patient age, gender, and BMI. Additionally, to control for variations in case complexity, groups A and B were matched by age, gender, BMI, lesion location, and intervention type. Propensity score matching and a paired t test were used to compare the KAP, RAK, and FT stratified by single intervention procedures.

Results: A total of 487 endovascular procedures were performed: 209 in group A and 278 in group B. A total of 111 single intervention procedures from each group were matched (1:1), with a mean age of 61 ± 8 years and a BMI of 26.5 ± 4 kg/m. The median KAP, RAK, and FT for group A were 28.8 Gy · cm (interquartile range [IQR], 24-34 Gy · cm), 146 mGy (IQR, 123-173 mGy), and 12 minutes (IQR, 10-14 minutes), respectively. The median KAP, RAK, and FT for group B were 18.3 Gy · cm (IQR, 16-22 Gy · cm), 71.2 mGy (IQR, 60-85 mGy), and 10.4 minutes (IQR, 9-12 minutes), respectively. The KAP, RAK, and FT were significantly decreased in group B by 24% (P = .005), 41% (P < .001), and 22% (P = .002), respectively, compared with the values for group A. Stratified by single intervention procedures, the KAP and RAK had decreased significantly in group B (36% [P = .002] and 51% [P < .001], respectively) compared with group A. The FT decrease of 13% in group B was not statistically significant (P = .20).

Conclusions: Use of the Clear VD11 PURE platform system (Siemens Healthineers) reduced the patient radiation dose by 51% during endovascular peripheral interventions. The similar FTs for the matched single intervention procedures before and after the upgrade indicated consistent case complexity and surgeon practice. This platform appears to be an effective system for lowering the radiation dose.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2022.03.865DOI Listing

Publication Analysis

Top Keywords

radiation dose
20
kap rak
20
single intervention
16
endovascular procedures
12
siemens healthineers
12
age gender
12
group
12
intervention procedures
12
technology system
8
patient radiation
8

Similar Publications

Purpose: To evaluate the effects of four-dimensional noise reduction filtering using a similarity algorithm (4D-SF) on the image quality and tumor visibility of low-dose dynamic computed tomography (CT) in evaluating breast cancer.

Materials And Methods: Thirty-four patients with 38 lesions who underwent low-dose dynamic breast CT and were pathologically diagnosed with breast cancer were enrolled. Dynamic CT images were reconstructed using iterative reconstruction alone or in combination with 4D-SF.

View Article and Find Full Text PDF

While radiation hazards induced by cone-beam computed tomography (CBCT) in image-guided radiotherapy (IGRT) can be reduced by sparse-view sampling, the image quality is inevitably degraded. We propose a deep learning-based multi-view projection synthesis (DLMPS) approach to improve the quality of sparse-view low-dose CBCT images. In the proposed DLMPS approach, linear interpolation was first applied to sparse-view projections and the projections were rearranged into sinograms; these sinograms were processed with a sinogram restoration model and then rearranged back into projections.

View Article and Find Full Text PDF

Radiopharmaceuticals targeting prostate-specific membrane antigen (PSMA) have emerged as a sensitive tool for PET imaging of prostate cancer (PCa) recurrence. Yet urinary bladder activity may obscure the visualization of prostate bed recurrence. Among the Food and Drug Administration-approved PSMA radiopharmaceuticals, F-flotufolastat (rhPSMA-7.

View Article and Find Full Text PDF

Background: Chordoma is a slow-growing, primary malignant bone tumor that arises from notochordal tissue in the midline of the axial skeleton. Surgical excision with negative margins is the mainstay of treatment, but high local recurrence rates are reported even with negative margins. High-dose radiation therapy (RT), such as with proton or carbon ions, has been used as an alternative to surgery, but late local failure remains a problem.

View Article and Find Full Text PDF

Ionizing radiation exposure during perinatal development can produce various biological effects on the developing offspring. These effects are dependent on a number of factors, including total dose, dose rate and the developmental processes occurring at the time of irradiation. The present study conducted an analysis of historical radiobiological archived data involving 60Co-gamma irradiation of beagle dogs at specific periods of prenatal or postnatal development.

View Article and Find Full Text PDF