Purpose: To provide plan backup resiliency for patients treated on a solitary high definition multileaf collimator (HDMLC) linac by developing a fully integrated Eclipse script, which converts patient plans initially optimized on Millennium-120 (M120) MLC to dosimetrically equivalent leaf motions for delivery on HDMLC. In the event of HDMLC machine downtime, affected patients can be transferred to Millennium-120 units, and their backup plan delivered without delay.
Methods: Write-enabled Eclipse scripting is leveraged to generate HDMLC treatment fields with control points parameterized to mimic apertures of an existing Millennium-120 VMAT plan. Non-parity between intermediate control point gantry angles of script generated arcs relative to VMAT is reconciled through an interpolation subroutine to correct for the apertures and monitor units that would have existed at intermediate angles. Differences in dosimetric leaf gap are corrected by displacing the subset of leaves undergoing dynamic motion. A nominal change to plan normalization corrects for remaining discrepancies between beam models.
Results: Over 220 non-SABR VMAT patients were treated on a solitary HDMLC linac with plans converted using the developed script. All have undergone streamlined RO review and physics quality assurance (QA), where the converted plan replicates the original leaf patterns, representing a minor dosimetric perturbation. Analyzing a subset of converted plans delivered at four anatomical sites, on average 99.3% of points pass the 1%/1 mm gamma criterion. Dose-volume histograms between the original and converted plans are in excellent agreement. ArcCheck measurements comparing delivery of the converted HDMLC plan to the calculated M120 dose distribution averaged a gamma pass rate of 99.4% (95.2%) at a 3%/3 mm (2%/2 mm) criterion. The conversion process takes 30 s to run, avoids errors in exporting/re-importing, and generates leaf motions deliverable within machine limits.
Conclusion: The methodology developed for automated plan conversion helped maximize the utilization of a solitary HDMLC linac, while preserving backup interoperability with minimal overhead.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195034 | PMC |
http://dx.doi.org/10.1002/acm2.13598 | DOI Listing |
J Appl Clin Med Phys
November 2024
Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Purpose: The aim was to develop and evaluate an EPID-based MLC positional test that addresses known weaknesses of the picket fence test and has sufficient accuracy so that the AAPM MPPG 8.b. MLC position action limit of ± 0.
View Article and Find Full Text PDFJ Appl Clin Med Phys
January 2025
Department of Radiation Oncology, University of Rochester Medical Center, Rochester, New York, USA.
Purpose: To compare the effect of two linacs designs on fetal dose sparing on a pregnant patient, including estimation of the fetal dose, and the effect of a lead apron.
Methods: A patient with a high-grade sarcoma located in the right knee/lower thigh was prescribed 51 Gy (1.7 Gy/Fx) with a simultaneous-integrated-boost (SIB) of 60 Gy to a smaller volume, starting in the 26th gestational week.
Biomed Phys Eng Express
August 2024
Department of Physics and Atmospheric Science, Dalhousie University, Halifax, Nova Scotia, Canada.
. This work investigates the small-field dosimetric characteristics of a 2.5 MV sintered diamond target beam and its feasibility for use in linac-based intracranial stereotactic treatments.
View Article and Find Full Text PDFOpen Vet J
July 2022
Joint faculty of Veterinary Medicine, Yamaguchi University, Yamaguchi-shi, Japan.
Background: For optimal treatment, it is important to maintain optimal multi-leaf collimator (MLC) transmission in intensity-modulated radiation therapy (IMRT). However, adjustment of transmissions has not been reported in veterinary medicine.
Aim: To demonstrate that appropriate MLC parameter adjustment for IMRT using 4- and 6-MV energy can reduce the need for quality assurance revalidation in real companion animal clinical cases.
J Appl Clin Med Phys
June 2022
Department of Medical Physics, BC Cancer - Centre for the Southern Interior, Kelowna, British Columbia, Canada.
Purpose: To provide plan backup resiliency for patients treated on a solitary high definition multileaf collimator (HDMLC) linac by developing a fully integrated Eclipse script, which converts patient plans initially optimized on Millennium-120 (M120) MLC to dosimetrically equivalent leaf motions for delivery on HDMLC. In the event of HDMLC machine downtime, affected patients can be transferred to Millennium-120 units, and their backup plan delivered without delay.
Methods: Write-enabled Eclipse scripting is leveraged to generate HDMLC treatment fields with control points parameterized to mimic apertures of an existing Millennium-120 VMAT plan.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!